<strong>Summit</strong>Electronic Analytical and Precision BalancesElektronische Analysen- undPräzisionswaagenOperating Instructions/Betriebsanleitung


English – page 3Deutsch – Seite 602


ContentsWarnings and Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Below-Balance Weighing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Leveling the Balance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Display and Operating Elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Basic Weighing Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Calibration/Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15Configuration (Operating Menu) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Functions of the Keys during Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Menu Navigation; Example: Setting the Language . . . . . . . . . . . . . . . . . . . . . . 19Parameter Settings: Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Parameter Settings: Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Input: ID Number, Date and Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Application Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Counting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Weighing in Percent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Animal Weighing/Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Net-total Formulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36Totalizing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Density Determination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40Mass Unit Conversion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42ISO/GLP-compliant Printout/Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45Data Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Troubleshooting Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49Recycling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56C Marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57Declaration of Type Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 583


Warnings and Safety PrecautionsSafety§ To prevent damage to the equipment,please read these operatinginstructions carefully before usingthe balance.! Do not use this equipment inhazardous areas.! The balance may be opened onlyby trained service technicians.! Disconnect the balance frompower before connecting ordisconnecting peripheral devices.! If you operate the balance underambient conditions subject tohigher safety standards, you mustcomply with the applicable installationregulations.! Exposure to excessive electromagneticinterference can causethe readout value to change.Once the disturbance has ceased,the instrument can be used againin accordance with its intendedpurpose.Make sure that no liquid entersthe equipment housing; use onlya slightly moistened cloth toclean the balance.Installation! Make sure the voltage rating printedon the power supply is identicalto your local line voltage.– Proceed with extreme cautionwhen using pre-wired RS-232connecting cables, as the pinassignments may not be compatiblewith <strong>Denver</strong> equipment.Before connecting the cable,check all pin assignments againstthe cabling diagrams and disconnectany lines that are assigned4 differently.! If there is visible damage to theequipment or power cord, disconnectthe equipment from powerand lock it in a secure place toensure that it cannot be used forthe time being.– Connect only <strong>Denver</strong> accessories,as these are optimally designedfor use with your <strong>Summit</strong> balance.The operator shall be solelyresponsible for installation andtesting of any modifications to<strong>Denver</strong> equipment, including connectionof cables or equipmentnot supplied by <strong>Denver</strong>. Onrequest, <strong>Denver</strong> will be happy toprovide information on operatingspecifications (in accordance withthe Standards for defined immunityto interference).$ Do not open the balance housing.If the seal is broken, this will voidall claims under the manufacturer’swarranty.$ If you have any problems withyour balance, contact your local<strong>Denver</strong> customer service center.SymbolsThe following symbols are usedin these instructions:§ indicates required steps$ indicates steps required onlyunder certain conditions> describes what happens after youhave performed a particular step– indicates an item in a list! indicates a hazard


Getting StartedStorage and Shipping Conditions– Do not expose the balance toextreme temperatures, moisture,shocks, blows or vibration.Unpacking the Equipment§ After unpacking the equipment,please check it immediately forany external damage.$ If you detect any damage, proceedas directed in the chapter entitled“Care and Maintenance,” under“Safety Inspection.”$ Save the box and all parts of thepackaging for any future transport.Disconnect all cables before packingthe balance for shipping.Equipment Supplied– Balance– Weighing pan– Pan support (only for modelswith a round weighing pan)– AC adapterAdditional equipment supplied withmodels S/SI-234, -114, -64:– Sliding-door draft shield– Drip/breeze ring– Draft shield base plateInstallationChoose a location that is not subjectto the following negative influences:– Heat (heater or direct sunlight)– Drafts from open windows anddoors– Excessive vibration duringweighing– Excessive moistureConditioning the BalanceMoisture in the air can condenseon cold surfaces whenever the equipmentis moved to a substantiallywarmer place. To avoid the effectsof condensation, allow the balance tosit for 2 hours, at room temperature,before plugging into AC power.Seal on Balances Verified for Use inLegal Metrology in the EU*:EU legislation requires that a controlseal be affixed to verified balances.The control seal consists of a sticker.If the seal is broken, the verificationbecomes null and void and thebalance must be re-verified.Additional equipment supplied withmodels S/SI-603, -403, -203:– Round glass draft shield with coverAdditional equipment supplied withmodel SI-203DS:– Sliding-door draft shield* Including the Signatories of theAgreement on the EuropeanEconomic Area5


InstallationSetting Up the Balance<strong>Instrument</strong>s with sliding-door draft shield:§ Place components inside the chamber in thefollowing order:– Draft shield base plate– Drip/breeze ring (not on model SI-203DS)– Pan support– Weighing pan<strong>Instrument</strong>s with a round glass draft ring:§ Position the components listed below in theorder given:– Place the lower lid on the balance with theraised edge facing upwards and turn it untilit is firmly in position– Pan support– Weighing pan– Glass draft ring– Place the upper lid on the draft shield ringwith the raised edge facing downwards<strong>Instrument</strong>s with a round weighing pan§ Position the components listed below in theorder given:– Pan support– Weighing pan<strong>Instrument</strong>s with a rectangular weighing pan:§ Place the weighing pan on the balance6


Connecting the Balance to AC Power/SafetyPrecautionsUse only original <strong>Denver</strong> AC adapters.For use within– Europe: part no. 6971412– U.S./Canada: part no. 6971413§ Connect the angle plug to the balance§ Connect the AC adapter to the wall outlet(mains)AC Adapter with Country-specific Power CordSome models come with separate countryspecificpower cords for the AC adapter.§ Connect the angle plug to the balance§ Select the power cord for your area andconnect it to the AC adapter§ Plug the power cord into the wall outlet(mains)§ Use an original <strong>Denver</strong> AC adapter witha wide input voltage range (100 to 240 V~),order no. 6971966, and replaceable powercord:6900900 (Europe) 6900905 (AUS)6900901 (US/CDN) 6900902 (ZA)6971945 (UK) 6971977 (Argentina)6971973 (India) 6971978 (China)6971980 (Denmark) 6971975 (Israel)6971776 (Italy)Safety PrecautionsPlug-in AC Adapter 6971412/6971413:The AC adapter rated to Class 2 can be pluggedinto any wall outlet without additional safetyprecautions.Benchtop AC Adapter 6971966:The AC adapter rated to Class 1 can beplugged into any wall outlet without additionalsafety precautions.The ground terminal is connected to thebalance housing, which can be additionallygrounded for operation. The data interface isalso electrically connected to the balancehousing (ground).7


NOTE: This equipment has beentested and found to comply withthe limits pursuant to part 15 of FCCRules. These limits are designed toprovide reasonable protectionagainst harmful interference. Thisequipment generates, uses and canradiate radio frequency energy and, ifnot installed and used in accordancewith these instructions, may causeharmful interference to radio communications.For information on the specific limitsand class of this equipment, pleaserefer to the Declaration of Conformity.Depending on the particular class,you are either required or requestedto correct the interference.If you have a Class A digital device,you need to comply with the FCCstatement as follows: “Operation ofthis equipment in a residential area islikely to cause harmful interference inwhich case the user will be requiredto correct the interference at his ownexpense.”If you have a Class B digital device,please read and follow the FCCinformation given below:“However, there is no guarantee thatinterference will not occur in a particularinstallation. If this equipmentdoes cause harmful interference toradio or television reception, whichcan be determined by turning theequipment off and on, the user isencouraged to try to correct the interferenceby one or more of the followingmeasures:– Reorient or relocate the receivingantenna.– Increase the separation betweenthe equipment and receiver.– Connect the equipment into anoutlet on a circuit different fromthat to which the receiver is connected.– Consult the dealer or an experiencedradio/TV technician forhelp.”Before you operate this equipment,check which FCC class (Class A orClass B) it has according to theDeclaration of Conformity included.Be sure to observe the informationof this Declaration.8


Connecting Electronic Peripheral Devices§ Make sure to unplug the balance from ACpower before you connect or disconnect aperipheral device (printer or computer) to orfrom the interface port.Warmup TimeTo ensure accurate results, the balance mustwarm up for 30 minutes before operation.Only after this time will the instrument havereached the required operating temperature.Using Verified Balances in Legal Metrology inthe EU*:$ Make sure to allow the equipment to warmup for at least 24 hours after initial connectionto AC power or after a relatively longpower outage.Operation Outside the Temperature Range!Operating the balance beyond the temperaturerange of +10…+30°C (50°…86°F).Differences from the specifications listed inthe chapter on “Specifications” are possible.* Including the Signatories of the Agreementon the European Economic Area9


Below-Balance WeighingA port for a below-balance weighing hangeris located on the bottom of the balance.$ Below-balance weighing is not permitted inlegal metrology.§ Open cover plate on the bottom of thebalance.Important: set the balance on its side toaccess the cover plate. DO NOT turn thebalance upside-down.§ Using the built-in hook 1: Attach the sample(e.g., using a suspension wire) to the hanger.21$ Bore hole 2 (not on models S/SI-203, -602,-402, -8001, -4001): Carefully fasten thespecial hanger, or order a hanger directlyfrom <strong>Denver</strong>.$ Install a shield for protection against draftsif necessary.10


Leveling the BalancePurpose:– To compensate for unevenness at the placeof installationAlways level the balance again any time afterit has been moved to a different location.Only the 2 front feet are adjusted to level thebalance.§ Retract the two rear feet (only on modelswith a rectangular weighing pan).§ Turn the 2 front feet as shown in the diagramuntil the air bubble is centered withinthe circle of the level indicator.> In most cases this will require severaladjustment steps.§ On models with a rectangular weighing pan:Lower the 2 rear feet until they touch thesurface on which the balance rests.Anti-theft Locking Device§ To secure the balance at the place of installation,fasten a chain or a lock to the luglocated on the rear panel of the balance.11


OperationOverview of Display Elements12121110987Pos. Designation1 Weight unit2 Menu level indicator3 Symbol: “GLP printing mode active”4 Symbol: “Printing mode active”5 Symbol: “Application program active”6 Calculated-value indicator(i.e., not a weight value)7 Symbol: Gross or net value8 Symbols for active application(W, Z, L, V, R, A, C)9 Symbol: Calibration/adjustment function10 Symbols for zero range3456Pos. Designation11 Busy symbol: command is being processed(for example, “Wait for stability icon);after you turn on the power, A will bedisplayed until you press a key12 Weight value displayed in selected weightunitSymbols: Next item on current menu level↵ Select a parameter settingOverview of Operating Elements871654Pos. Designation1 Taring2 Start an application program3 Select an application program | Open theoperating menu4 ClearThis key is generally used to cancel functions:– Quit application program– Cancel calibration/adjustment routine |Exit the operating menuPos. Designation5 Start calibration/adjustment routine6 Data output:Press this key to send readout values tothe built-in data interface.7 On/off8 Level indicator2312


Basic Weighing FunctionPreparation§ Switch on the balance: Press§ Tare the balance, if necessary:Press t$ If necessary, change the configurationsettings:See the chapter entitled“Configuration”$ If desired, load the factory settings:See the chapter entitled“Configuration”Additional Functions$ Switching off the balance:Press$ Balance in standby mode:the current time is displayedFeatures– Taring the balance– Printing weightsUsing Verified Balances as LegalMeasuring <strong>Instrument</strong>s in the EU*:The type-approval certificate forverification applies only to non-automaticweighing instruments.For automatic operation with or withoutauxiliary measuring devices, youmust comply with the regulationsapplicable to the place of installation.§ Before using the balance as a legalmeasuring instrument, calibrateand adjust it at the place of useusing the built-in motorized calibrationweight; for details, see“Calibration/Adjustment” in thischapter.$ The temperature range (°C)indicated on the verification labelmust not be exceeded duringoperation.Example:BD ED 200K +10°C to +30°C* Including the Signatories ofthe Agreement on the EruopeanEconomic Area13


ExampleSimple WeighingStep Key (or instruction) Display/PrintoutBalance in standby mode11:54 AM1. Switch on the balance 0.0 gSelf-test is performed,followed by automaticinitial tare function.2. Place container on weighing pan + 11.5 g(in this example: 11.5 g).3. Tare the balance t 0.0 g4. Place sample in container + 132.0 g(in this example: 132 g).5. Print weight. PRINTN + 132.0 g14


Calibration and AdjustmentPurposeCalibration is the determination ofany difference between the measuredvalue displayed and the true weight(mass) of a sample. Adjustment is thecorrection of this difference, or itsreduction to an allowable level withinmaximum permissible error limits.To block calibration/adjustment:– Select Cal.-Adj.: blocked in themenu– Close the menu access switch onthe back of the balanceUsing Verified Balancess as LegalMeasuring <strong>Instrument</strong>s in the EU*:Before using your balance as a legalmeasuring instrument, internal calibrationmust be performed at theplace of installation.FeaturesCalibration/adjustment can beperformed only when:– there is no load on the balance– the balance is tared– the internal signal is stable– for external calibration, the valuedisplayed for the calibration weighton the balance does not differ fromthe nominal weight value by morethan 2%If these conditions are not met, anerror message is displayed (“Err 02”).You can use any of the followingweight units in calibration/adjustment:Cal.Unit: Grams, Kilogr. or Pounds(not for verified models)External Calibration in VerifiedBalancesWhen the balance is used in legalmetrology, external calibration isblocked as follows:– The setting of the menu accessswitch is locked (see “To blockcalibration/adjustment”)– The cap over the menu accessswitch is sealedFor details on generating an ISO/GLPcompliantprintout of calibration/adjustment results, see page 45.Following calibration/adjustment,the application program is cleared.Internal Calibration/AdjustmentIn the operating menu, select Cal.-Adj.: Cal.Int. before beginning.The built-in motorized calibrationweight is applied and removed automaticallyfor internal calibration.§ Select calibration/adjustment:Press q> The built-in weight is applied automatically> The balance is adjusted> The built-in calibration weight isremoved.* Including the Signatories of the Agreementon the European Economic Area15


Internal Calibration/Adjustment (Only on SI models)Set the following parameters:Setup: Bal.Scal.: Cal.-Adj.: Cal.Int. (menu code 1.1. 9. 4)The built-in motorized calibration weight is applied and removed automaticallyfor internal calibration.Step Key (or instruction) Display1. Tare the balance t 0.0 g2. Start calibration q Cal.Int.The built-in weight isapplied automaticallyCal. run.3. Calibration/adjustment executed Cal.end4. The built-in weight is removed 0.0 g16


External CalibrationParameters (changes in factory settings):SETUP: Bal.Scal.: Cal.-Adj.: Cal.Ext. (menu code 1.1. 9. 1)The required calibration weight is configured at the factory (see “Specifications”)Step Key (or instruction) Display1. Tare the balance t 0.0 g2. Start calibration. q Cal.Ext.Once you store the zero pointthe required calibration weightis prompted (flashing display)- 5000.0 g3. Apply the prompted calibration 5000.0 gweight (in this example: 5000 g)Weight too light:a minus sign “–” is shownWeight too heavy:a plus sign “+” is shownThe display stops flashingas soon as the weight value iswithin the defined limit.4. Calibration/adjustment executed; Cal.endthen the calibration weightis displayed+ 5000.0 g5. Remove the calibration weight 0.0 g17


Configuration (Operating Menu)You can configure the balance; i.e., adapt it to individual requirements.Functions of the Keys during ConfigurationSymbol Key FunctionV SELECT Press and hold Scroll through menu items> ENTEROne menu level lower↵ENTERConfirm menu itemCLEAR Press and hold Save settings and exit menu from anyposition


Menu NavigationExample: Setting the LanguageStep Key (or instruction) Display1. Open the menu: SELECT (hold) Applic.In weighing mode: first menu itemis shown2. Scroll upward within the Repeatedly: Inputmenu level; after the last SELECT...menu code, the firstcode is displayedagainlanguag.3. Select menu level ENTEREnglisho(scrolls to the right)4. Change setting: SELECTEspanolScroll until the desiredsetting is shown5. Confirm the menu code; ENTEREspanol o“o” indicates the active setting6. Return to the next higher menu CLEARLengualevel (from the second level)$ Set other menu items as desired7. Save settings Repeatedly:and exit menuCLEARor$ Exit menu without savingchanges> Restart your application 0.0 g19


Parameter Settings: MenuLevel 1 Level 2 Level 3 Menu codeSetup Bal.Scal. Ambient Ambient conditions 1. 1. 1.balance parameters App.filt. Application filter 1. 1. 2.Stab.Rng. Stability range 1. 1. 3.Taring Taring 1 ) 1. 1. 5AutoZer. Auto zero 1. 1. 6Wt.Unit Basic weight unit 1. 1. 7.Display Display accuracy 1 ) 1. 1. 8.Cal./adj. Function of the q key 1. 1. 9.Cal.Unit Weight unit for calibration 1 ) 1. 1. 11.Interf. Interface Baud Baud rate 1. 5. 1.Parity Parity 1. 5. 2.StopBit Number of stop bits 1. 5. 3.Handshk Handshake mode 1. 5. 4.Databit Number of data bits 1. 5. 5.Dat.Rec. Output: SBI (ASCII) or printout 1. 5. 6.Prnt.Out Print (manual/automatic) 1. 6. 1.Settings for print function StopAut. Stop automatic printing 1. 6. 2.Tar./prt. Tare balance after ind. print 1. 6. 4.Prt.Init. Printout of appl. parameters 1. 6. 5.Format Line format for printout 1. 6. 6.GLP ISO/GLP-compliant printout 1. 6. 7.time: 12/24 h 1. 6. 8.date: Format 1. 6. 9.Extras Menu 1. 8. 1.Additional functions Signal Acoustic signal (beep) 1. 8. 2.Keys Keypad 1. 8. 3.On Mode Power-on mode 1. 8. 5.Backlit Display backlighting 1. 8. 6.Reset Menu Factory settings 1. 9. 1.Applic. Weigh 2. 1.Application Unit Toggle wt. unit Disp.Dig. Display accuracy 1 ) 2. 2. 2.programs Count. Counting Resolut. Resolution 2. 3. 1.Ref.Updt. Autom. ref. sample updating 2. 3. 2.Percent Weighing in percent Dec.Plcs Decimal places 2. 4. 1.Net-Tot Net-total formulation Comp.Prt. Printout of components 2. 5. 1.Total Totalizing Comp.Prt. Printout of components 2. 6. 1.AnimalW. Animal weighing Activty. Animal activity 2. 7. 1.Start 2. 7. 2.Calc. Calculation Method (operator) 2. 8. 1.Dec.Plcs Decimal places 2. 8. 2.Density Density determination Dec.Plcs Decimal places 2. 9. 1.Input Input ID no., Date, time Input: ID no., date, time 3. 1./2./3.Info Information Version , Ser. No. , Model Display software ver., serial no., model 4. 1./2./3.Languag. English (factory setting) 5. 1.Deutsch German 5. 2.Franc. French 5. 3.Ital. Italian 5. 4.Espanol Spanish 5. 5.Codes Menu shows codes (not texts) 5. 6.1) Setting cannot be changed on verified balances20


Parameter Settings: Overviewο = Factory setting √ = User-defined settingLevel 1 Level 2 Level 3 Level 4 Menu codeSetup Bal.Scal. Ambient V.Stable Very stable 1. 1. 1. 1balance Ambient ο Stable 1. 1. 1. 2parameters conditions Unstabl 1. 1. 1. 3(Filter adaptation) V.Unstbl. Very unstable 1. 1. 1. 41 ) Setting cannot be changed on verified balances2 ) Only on SI modelsApp.filt. ο Final.Rd. Final readout mode 1. 1. 2. 1Application filter Filling Filling mode 1. 1. 2. 2Stab.Rng. 1/4 Dig. (digit) 1. 1. 3. 1Stability range 1/2 Dig. 1. 1. 3. 21 - Digit 1. 1. 3. 3ο 2 - Digit 1. 1. 3. 44 - Digit 1. 1. 3. 58 - Digit 1 ) 1. 1. 3. 6Taring 1 ) W/o Stb W/o stability 1. 1. 5. 1Taring ο W/ StAb After stability 1. 1. 5. 2AutoZer. Off 1. 1. 6. 1Auto zero ο On 1. 1. 6. 2Wt.Unit For list of units, see 1. 1. 7. 1Basic weight“Toggling between Weightunit Units” 1. 1. 7. 23Disp.Dig. 1 ) ο All 1. 1. 8. 1Display Minus 1 1. 1. 8. 2accuracyCal./adj. ο Cal. ext. External cal./adj. 1 ) 1. 1. 9. 1Function of the Cal. int. Internal cal./adj. 2 ) 1. 1. 9. 2q key Blocked q key blocked 1. 1. 3. 3Cal.Unit. Unit 1 ) ο Grams 1. 1.11. 1for calibration Kilogr. Kilograms 1. 1.11. 2weight Pounds 1. 1.11. 321


Level 1 Level 2 Level 3 Level 4 Menu codeSetup Interf. Baud rate 600 1. 5. 1. 3Interface ο 1200 1. 5. 1. 42400 1. 5. 1. 54800 1. 5. 1. 69600 1. 5. 1. 719200 1. 5. 1. 8Parity ο Odd 1. 5. 2. 3Parity Even 1. 5. 2. 4None 1. 5. 2. 5StopBit ο 1 Bit 1. 5. 3. 1No. of stop bits 2 Bits 1. 5. 3. 2Handshk. Sftware 1. 5. 4. 1Handshake ο Hrdware 1. 5. 4. 2mode None 1. 5. 4. 3DataBit ο 7 Bits 1. 5. 5. 1No. of data bits 8 Bits 1. 5. 5. 2Dat.Rec. Com- SBI (ASCII) 1 ) 1. 5. 6. 1munication mode ο Printer (GLP-printout) 1. 5. 6. 2Prnt.Out Print Man. W/o W/o stability 1. 6. 1. 1Printing fct. (manual/ ο Man.With W/ stability 1. 6. 1. 2automatic) Aut. W/o Autom. w/o stability 1. 6. 1. 3Aut.With. Autom. w/ stability 1. 6. 1. 4StopAut. Stop ο Off Not possible 1. 6. 2. 1automatic printing On Use print key PRINT 1. 6. 2. 2Tar./prt. Tare ο off 1. 6. 4. 1the balance after On 1. 6. 4. 2individual printout1) Note concerning verified balances as legal measuring instruments in the EU*:In the setting “SBI”, the non-verified display digit is not automatically identified.Please take the corresponding measures or adjust the settings on the peripheral device.* Including the signatories of the Agreement on the European Economic Area.22


Level 1 Level 2 Level 3 Level 4 Menu codeSetup Prnt.Out Prt.Init. Off 1. 6. 5. 1Printing Printing appli- ο All All parameters 1. 6. 5. 2cation parameters MainPar. Main parameters 1. 6. 5. 2Format Line 16 Char. 16 characters (w/o ID) 1. 6. 6. 1format for printout ο 22 Char. 22 characters (w/ ID) 1. 6. 6. 22 Char. 2 characters (w/ ID) 1. 6. 6. 3(Date/Time and weight value)GLP Printout ο Off 1. 6. 7. 1as ISO/GLP- Cal.-adj. Only for calib./adj. 1. 6. 7. 2compliant Always All printouts 1. 6. 7. 3time 24 h 24-hour format 1. 6. 8. 112 h 12-hour format “AM/PM” 1. 6. 8. 2date dd.mmm.yy Day/month/year 1. 6. 9. 1mmm.dd.yy Month/day/year 1. 6. 9. 2Extras Menu ο CanEdit Can change settings 1. 8. 1. 1Additional Rd. Only Read only 1. 8. 1. 2functionsSignal Off 1. 8. 2. 1Acoustic signal ο On 1. 8. 2. 2Keys ο Free 1. 8. 3. 1Keypad Locked 1. 8. 3. 2On Mode ο Off/On Off/on/standby 1. 8. 5. 1Power-on Standby On/standby 1. 8. 5. 2mode Auto On Auto on 1. 8. 5. 3Backlit Off 1. 8. 6. 1Display ο On 1. 8. 6. 2backlightingReset Menu Yes Restore defaults 1. 9. 1. 1Reset menu Factory settings ο No Do not restore defaults 1. 9. 1. 223


Level 1 Level 2 Level 3 Level 4 Menu codeApplic. Weigh 2. 1.Applic. Unit Disp.Dig. 1 ) ο All 2. 2. 2. 1programs Toggle units Display Minus 1 2. 2. 2. 2accuracyCounting Resolut. ο Disp.Acc. Display accuracy 2. 3. 1. 1Resolution 10-fold 10 times > disp. 2. 3. 1. 2Ref.Updt. ο Off 2. 3. 2. 1Autom. reference Auto 2. 3. 2. 2updatingPercent Dec.Plcs None No dec. places 2. 4. 1. 1Weighing Decimal places ο 1 Dec.Pl. 1 decimal place 2. 4. 1. 2in percent 2 Dec.Pl. 2 decimal places 2. 4. 1. 33 Dec.Pl. 3 decimal places 2. 4. 1. 4Net-Tot Comp.Prt. Off 2. 5. 1. 1Net-total Component ο On 2. 5. 1. 2printoutTotal Comp.Prt. Off 2. 6. 1. 1Totalizing Component ο On 2. 6. 1. 2printoutAnimalW. Activty. Calm (fluct.: 2% of test obj.) 2. 7. 1. 1Animal Animal activity ο Active (fluct.: 5% of test obj.) 2. 7. 1. 2weighing V.Active(fluct.: 20% of test obj.) 2. 7. 1. 3Start Manual 2. 7. 2. 1ο Auto. Automatic 2. 7. 2. 2Calc. Method ο Mul. Multiplier 2. 8. 1. 1Calculation (operator) Div. Divisor 2. 8. 1. 2Dec.Plcs None No dec. places 2. 8. 2. 1Decimal ο 1 Dec.Pl.1 decimal place 2. 8. 2. 2places 2 Dec.Pl.2 decimal places 2. 8. 2. 33 Dec. Pl. 3 decimal places 2. 8. 2. 4Density Dec.Plcs None No dec. places 2. 9. 1. 1Density Decimal places ο 1 Dec.Pl.1 decimal place 2. 9. 1. 2determination1 ) Setting cannot be changed on verified balances24


Input: ID number, Date and TimeLevel 1 Level 2 Level 3 MenuCodeInput ID No. ID input for ISO/GLP compliant data record; 3. 1.Input7 characters max.Permitted characters: 0 to 9; A to Z;dash/hyphen; spaceDATE Menu item for setting the date 3. 2.TIME Menu item for setting the time 3. 3.Depending on the setting you select for the menu item “SETUP PRNT.OUT DATE,”the date will be displayed in the following format:FormatDD.MMM.YYDisplay: DateMMM.DD.YYDepending on the setting you select for the menu item “SETUP PRNT.OUT TIME,”the time will be displayed in the following format:Time24-hour formatDisplay: Time12-hour format25


Example: ID No., Date and TimeStep Key (or instruction) Display1. Open the menu: SELECT hold APPLIC.In weighing mode; firstmenu item is displayed2. Select “Input” SELECTINPUT3. Select input for ID no. ENTER twice ID NO.4. Set or change the ID no. – SELECT3------hold down key to automaticallychange the digit(s):SELECT hold5. Scroll within the 7-digit ENTER or 3-aBC12ID no.6. Save input when youCLEARENTERID NO.have reached the last digitof the ID no.7. Select “Date” SELECT ENTER, FEB. 08.8. Change setting – SELECTFEB. 10.hold down key to automaticallychange the digit(s):SELECT hold9. Toggle between ENTER or FEB. 10.Day/Month/Year positions10. Save setting when you reachthe “YEAR” positionCLEARENTERdate11. Select “Time” SELECT ENTER, A10.46.2312. Change setting –hold down key to automatically SELECTchange the digit(s): SELECT hold A11.46.2313. Toggle between ENTER or A11.46.32Hour/Minute/Second positions14. Set seconds to zeroCLEARSELECTA11.47.0015. Save setting when you havereached the “second” positionENTERTIME16. Save all settingsCLEARand exit the menurepeatedly> Restart your application 0.0 g26


Device InformationLevel 1 Level 2 Level 3 Example MenucodeInFo version Show software version rel.36.01 4. 1.InformationSer. No. Show serial number(To toggle focus between10801234 4. 2.upper and lower displaysections, press SELECT )Model Show model designation SI-8001 4. 3.(to change focus from upperto middle to lower displaysection and back, pressSELECT )Display of Menu Items: Text or CodesLanguag. English (factory setting) 5. 1.Deutsch German 5. 2.Franc. French 5. 3.Ital. Italian 5. 4.Espanol Spanish 5. 5.Codes Menu shows codes (not texts) 5. 6.Application ProgramsUsing Verified Balances as Legal Measuring <strong>Instrument</strong>s in the EU*:All application programs can be selected on balances used as legal measuringinstruments.Calculated values are alternately indicated with the following symbols:– Percent = %– Piece count (Counting) = pcs– Computed value = o,G* Including the Signatories of the Agreement on the European Economic Area27


CountingDisplay symbol: ZPurposeWith the Counting program you candetermine the number of parts thateach have approximately equal weight.To do this, a known number of parts(the reference sample quantity) isweighed first, and the individualpiece weight (reference weight) iscalculated from this result.Thus the number of parts subsequentlyplaced on the balance can bedetermined from their weight.Preparation§ Select the Counting application inthe menu:see “Configuration.”§ Set the following parameters:Applic. Application programCount.Resolut. Resolutiono Disp.Acc. Display accuracy10-fold 10-fold higherRef.Updt. Autom. ref. sampleupdatingo OffAutom.o = Factory settingDisplay accuracyAutomaticChanging the Reference SampleQuantityActivate function:Press the SELECT keySelect the desired reference samplequantity (1 to 100):In increments of 1: Press the SELECT keybrieflyIn increments of 10:Press and hold the SELECT key.The quantity is stored in batterybackedmemory.Reference Sample UpdatingAutomatic reference sample updatingoptimizes the counting accuracy.You can activate or deactivate thisfunction in the menu.Automatic reference sample updatingis performed when the requirements,including the specified stability criterion,have been met.The abbreviation opt, for “optimizing”,is displayed briefly with thenew reference sample quantity.Printout: CountingnRef + 10 : Reference samplequantitywRef + 21.14 g : Reference weightQnt + 500 pcs : Calculated quantity28


Example : Counting parts of equal weightParameter: Applic. - Count. (menu code 2. 3.)Step Key Display/Data output(or instruction)1. Place empty container + 22.6 gon the balance2. Tare the balance t 0.0 g3. Add reference sample quantityto container (in this example:20 pcs)4. Changing the reference SELECTref 10 pcssample quantity:5. Select reference sample quantity: Repeatedly: SELECTIn increments of 1 (1, 2, 3, etc. to 100) Press brieflyIn increments of 10 (10, 20, etc. to 100) SELECT pressand hold ref 20 pcs6. Confirm selected reference ENTER+ 20 pcssample quantity and start nRef*20 pcsapplication. The current reference wRef 1.07 gweight remains stored until a newreference weight remains storeduntil a new reference is set orthe power supply is interrupted7. Add desired number of pieces. + 500 pcs8. If desired, print quantity PRINTQnt + 500 pcs9. Toggle display between 1.07 g Gmean piece weight, weight, Repeatedly: + 535.0 g *quantity SELECT+ 500 * pcs*10. Unload the balance – 21 pcs *11. Repeat as needed, starting fromStep 712. Delete reference value CLEAR0.0 g29


Weighing in PercentDisplay symbol: %PurposeThis application program allows youto obtain weight readouts in percentwhich are in proportion to a referenceweight.Preparation§ Select the Weighing in percentapplication in the menu:see “Configuration.”Changing the Reference PercentageActivate function:Press the SELECT keySelect the desired reference (1 to100):In increments of 1: Press the SELECT keybrieflyIn increments of 10: Press and holdthe SELECT key.The percentage is stored in batterybackedmemory.§ Set the following parameters:Applic. Application programPercent Weighing in percentDec.Plcs. Decimal placesNone No decimal placeso 1 Dec. Pl. 1 decimal place2 Dec. Pl. 2 decimal places3 Dec. Pl. 3 decimal placeso = Factory settingPrintout: Weighing in percentpRef 100 : ReferencepercentageWxx% 111.6 g : Reference weightnet xx% forselected referencepercentagePrc + 94.9 % : Calculated referencepercentage30


Example: Determining residual weight in percentParameter settings: Applic.: Percent (menu code 2. 4.)Reference percentage: Ref 100%Step Key Display/Data output(or instruction)1. Tare the balance t 0.0 g2. Place sample equal to 100%on the balance(in this example: 111.6 g)3. Information:Enter reference percentage SELECTref 100 %(Changing the reference:see the previous page)4. Initialize the balance ENTER+ 100.0 %The current reference weight pRef*100 %remains stored until a new Wxx% 111.6 greference is set or the powersupply is interrupted5. Remove sample(e.g., for drying)6. Place unknown weight + 94.9 % *on balance(in this example: 105.9 g)7. If desired, print percentage PRINTPrc + 94.9 %8. Toggle display between Repeatedly: + 105.9 gweight and percentage SELECT+ 94.9 * %*9. Clear display of residual weightand reference percentage CLEAR+ 105.9 g10. If desired, print net PRINTN + 105.9 gresidual weight31


CalculationDisplay symbol: CPurposeWith this application program youcan calculate weight value using amultiplier or divisor. This can be used,for example, to determine the weightper unit area, or “gsm” weight (gramsper square meter), of paper.Preparation§ Select the Calculation application inthe menu:see “Configuration.”Setting the Factor or DivisorActivate function:Press the SELECT keySelect a number of up to 7 digits and,if needed, one decimal point(0.000001 to 9999999):In increments of 1: Press the SELECT keybrieflyTo increase the value without pressingrepeatedly:Press and hold the SELECT key.The selected operator is stored inbattery-backed memory.§ Set the following parameters:Applic. Application programCalc. CalculationMethod Operatoro Mul. MultiplierDiv. DivisorDec.Plcs. Decimal placesNone No decimal placeso 1 Dec. Pl. 1 decimal place2 Dec. Pl. 2 decimal places3 Dec. Pl. 3 decimal placesPrintout: CalculationMul + 1.2634 : MultiplierDiv + 0.6237 : DivisorRes + 79.7 o : Resulto = Factory setting32


Example:Calculating the weight per unit area of paper: An A4 sheet of paper is used inthis example, with surface dimensions of 0.210 m + 0.297 m = 0.06237 m 2 . Todetermine the weight per unit area, the total weight is divided by the surface.Parameter settings:Applic.: Calc..: Method: Div. (menu code 2. 8. 1. 2)Step Key Display/Data output(or instruction)1. Tare the balance t 0.00 g2. Activate divisor input SELECT______0.3. Set the divisor (in this example: 0.06237):Position the decimal point, ENTER ,Enter numeralsRepeatedly5+ SELECT , __.000002+ENTER__.06000or press and hold:SELECT ,ENTER , etc. __.062374. Store the divisor andinitialize the balance ENTER+ 0.0 oThe current divisor remains Div 0.6237stored in battery-backed memoryuntil the setting is changed5. Weight per unit area: + 79.7 o *Place an A4 sheet of paperon the balance6. If desired, print result PRINTRes + 79.7 o7. Toggle display between Repeatedly: + 4.97 gweight and calculated value SELECT+ 79.7 * o*8. Unload the balance + 0.0 o *9. Repeat as needed, starting fromStep 533


Animal Weighing/AveragingDisplay symbol: VPurposeUse this program to determine theweights of unstable samples (e.g., liveanimals) or to determine weightsunder unstable ambient conditions.With this program, the balance calculatesthe weight as the average of adefined number of individual weighingoperations (also referred to as “subweighingoperations”).Preparation§ Select the Animal weighingapplication in the menu:see “Configuration.”§ Set the following parameters:Applic. Application programAnimalW. Animal weighingactivty. Animal activityStartCalm Stable readouto Active UnstableV.Active Very unstable readoutmanualo Autom. AutomaticChanging the Number of SubweighingOperationsActivate function:Press the SELECT keySelect the desired number ofmeasurement (1 to 100):In increments of 1:Press the SELECT key brieflyIn increments of 10:Press and hold the SELECT key.The selected number of measurementsis stored in battery-backedmemory.Printout: Animal weighingmDef 20 : Number of subweighingoperationsx-Net + 410.1 g : Calculated averageo = Factory setting34


Example: Determining animal weight with automatic start and 20 subweighingoperations (measurements)Parameter settings: Applic.: AnimalW. (menu code 2. 7.)Step Key Display/Data output(or instruction)1. Place animal weighing bowl 22.6 gon the balance2. Tare the balance t 0.0 g3. Change the numberof subweighing operations: SELECTref 304. Set number of measurements: Repeatedly:In increments of 1 (1, 2, 3, etc. to 100) SELECT Press brieflyIn increments of 10 (10, 20, etc. to 100) SELECT press ref 20and hold5. Confirm number of measurements ENTERand start automatic animal weighing+ 0.0 g *The number of measurementsremains stored in battery-backedmemory until the setting is changed6. Place first animal in bowl. 888The balance delays the start 20of measurements until the 19difference between 2 measure- ....ments meets the criterion 1$ If you selected the manual startmode, pressENTER7. Read off the resultThe result is displayed with the+ 410.1 g G*“*” symbol (= calculated value) mDef 20and remains displayed until x-Net + 410.1 gthe sample (animal) is removedfrom the load plate (bowl)8. Unload the balance + 0.0 g *9. Weigh next animal (if desired)Next weighing series begins automatically35


Net-total FormulationDisplay symbol: RPurposeWith this application program you canweigh in individual components eitherby their individual weight or by thetotal weight. You can print out boththe total weight and the individualweights of the components.Preparation§ Select the Net-total application inthe menu:see “Configuration.”§ Set the following parameters:Applic. Application programNet-TotL. Net-total formulationComp.Prt. Printout of componentsOffo OnFeatures– Weigh up to 99 components from“0” to a defined total componentweight.– Store component weights(printout shows Comp xx), with– display zeroed automaticallyafter value is stored, and– automatic printout– Clear component memory followingcancellation of the weighingsequence (by pressing CLEAR ) andprintout of the total weight.– Toggling between componentweight and total weight by pressingand holding SELECT (< 2 sec).– Printout of the total of the individualcomponent weights (T-Comp)o = Factory settingPrintout: Net-total formulationComp 2+ 278.1 g : SecondcomponentT-Comp+ 2117.5 g : Sum ofcomponents36


Example: Counting parts into a containerParameter settings:Applic.: Net-Tot (menu code 2. 5.)Step Key (or instruction) Display/Data output1. Place empty containeron the balance.65.0 g2. Tare the balance t 0.0 g3. Add first component + 120.5 gENTER4. Store component data + 0.0 g *NETComp 1+ 120.5 g5. Add next component + 70.5 g *NET6. Store component data ENTER+ 0.0 g *NET7. Weigh in further components Repeat stepsas desired 5 and 6Comp 2+8. Fill to desired final value view SELECT+ 191.0 gthe current total weight value:*70.5 g9. Print total weight and clear CLEAR+ 2117.5 gthe component memory T-Comp+ 2117.5 g37


TotalizingDisplay symbol: RPurposeWith this application program youcan add values from successive,mutually independent weight valuesto a total that exceeds the capacity ofthe balance.Preparation§ Select the Totalizing application inthe menu:see “Configuration.”§ Set the following parameters:Applic. Application programTotal TotalizingComp.Prt. Printout of componentsOffo Ono = Factory settingFeatures– Totalizing memory for up to99 values– Store component weights (printoutshows Comp xx), with automaticprintout– Toggle display between the currentindividual weight value and thevalue in totalizing memory bypressing SELECT– Printout of the total of the individualcomponent weights (S-Comp)– To close the application programand print the total weight: pressCLEARPrintout: TotalizingComp 2+ 278.1 g : Second componentS-Comp+ 2117.5 g : Totalizing memory38


Example: Totalizing weight valuesParameter settings:Applic.: Total: Comp.Prt: On (menu code 2. 6. 1. 2)Step Key Display/Data output(or instruction)1. Tare the balance t 0.0 g2. Place sample balance + 380.0 g(in this example: 380 g)3. Store value in memory ENTER+ 380.0 g *Comp 1+4. Remove sample + 0.0 g *380.0 g5. Place the next sample on thebalance (in this example, 575 g) + 575.0 g *6. Store value in memory ENTER+ 955.0 g *+ 575.0 gComp 2+ *575.0 g7. View the value in totalizingmemory SELECT+ 955.0 g G*8. Weigh in further components Repeat stepsas desired 5 and 69. Print total weight and CLEAR0.0 gclear the totalizing memory S-Comp+ 2117.5 g39


Density DeterminationDisplay symbol: WPurposeThis application program lets youdetermine the density of solid substancesusing the buoyancy method.You can have results displayed withone decimal place, or no decimalplaces: see “Configuration.”Note: the sample holder and suspensionwire used in the example beloware not included with the balance.Printout for Density DeterminationWa + 20.0 g : Weight in airWfl + 15.0 g : Weight in liquidRho 4.0 o : Result: densityof the samplePreparation§ Select the Density Determinationapplication in the menu:see “Configuration.”§ Set the following parameters:Applic. Application programdensity Density determinationdec.plcs. Decimal placesnone No decimal placeso 1dec.pl. 1 decimal placeo = Factory setting40


Example: Determining the density of a solid sample.Parameter settings:Applic.: Density: Dec.Plcs: 1 Dec. Pl. (menu code 2. 9. 1. 2)Step Key Display/Data output(or instruction)1. Attach sample holder to suspension wire2. Tare the balance t 0.0 g3. Start application program4. Confirm “Air” displayENTERENTERAir ?5. Determine the weightof the sample in air:Place sample on the balance + 20.0 g *6. Store value for weight in air ENTER7. Remove sample from the balance water ?8. Determine weight in liquid:place sample in holder9. Confirm “water” display ENTER0.0 g *10. Immerse sample in liquid + 15.0 g *11. Store value for weight inENTERliquid, view result, and print + 4.0 o *Wa + 20.0 g12. Delete resultCLEAR13. Repeat as desired, starting from Step 3.Wfl + 15.0 gRho4.0 o41


Mass Unit ConversionPurposeWith this application program youcan change the weight value displayedfrom the basic weight unit toany of 4 application weight units (seetable on next page).Preparation§ Select the Unit application fortoggling weight units:see chapter on “Configuration”(Parameter Settings)Features– Set the basic unit and display accuracyin the Setup menu: see “Configuration.”– Set the application weight unitsand display accuracies in theApplication menu.– These settings are stored in battery-backedmemory.– The basic unit is active when thebalance is powered up.§ Set the following parameters:Application programunit Toggle wt, unitdec.plcs. Display accuracyo All All decimal placesminus 1 Reduced by 1 placeo = Factory settingPrintout for Wt. Unit Toggling+ 100.0 g : Weight with 16-character data output formatN + 0.22046 lb : Weight with 22-character data output format13-Jan-2005 08:35 : Data output format for two-line printout:N + 3.5275 ozt : Date/time and weight42


Example: Change display from the basic unit (in this example, grams [g]) topounds [lb] and then to Troy ounces [ozt].Set the following parameters: Applic.: Unit (code 2. 2.)Step Key (or instruction) DisplayPreparation:1. Begin selection of anapplication weight unit SELECTNone o2. Select an application unit;in this example, pounds(see table on next page)Repeatedly:Pounds3. Confirm the weight unit (pounds) ENTERpounds o4. Select the next applicationweight unit; ENTER , None oin this example: Troy ounces Repeatedly:(see table on next page)SELECTTroy Oz.SELECT5. Confirm weight unit (Troy ounces)ENTERTroy Oz. o6. Select other application units ifdesired (max. 4 total) (otherwise,confirm None by pressing ENTER )7. Save selection CLEAR0.00 gConversion:8. Place sample on balance + 100.00 g9. Toggle unit for weight value Repeatedly: + 0.22046 lbENTER+ 3.5275 ozt43


The following weight units are available in your <strong>Summit</strong> balance (in legalmetrology, only units permitted by national law are available):Menu item Unit Conversion Displayfactorsymbol1) UserDef. 1 ) Grams 1,00000000000 o2) Grams (factory setting) Grams 1.00000000000 g3) Kilogr. Kilograms 0.00100000000 kg4) Carats Carats 5.00000000000 o5) Pounds Pounds 0.00220462260 lb6) Ounces Ounces 0.03527396200 oz7) Troy Oz. Troy ounces 0.03215074700 ozt8) HKTael Hong Kong taels 0.02671725000 tl9) Sng.Tael. Singapore taels 0.02645544638 tl10) Twn.Tael Taiwanese taels 0.02666666000 tl11) Grains Grains 15.4323583500 GN12) Peny.Wt. Pennyweights 0.64301493100 dwt13) Milligr. Milligrams 1000.00000000 mg14) Pt.P.Lb. Parts per pound 1.12876677120 o15) Chn.Tael Chinese taels 0.02645547175 tl16) Mommes mommes 0.26670000000 m17) Austr.Ct. Austrian carats 5.00000000000 Kt18) Tola Tola 0.08573333810 o19) Baht Baht 0.06578947436 b20) Mesghal Mesghal 0.21700000000 o21) Tons Tons 0.00000100000 t22) Lb / Oz 2 ) Pounds : ounces 0.03527396200 lb oz23) Newton Newton 0.00980665000 N1) = User-defined weight unit; can be loaded in the balance over an optionalRS-232 or USB interface using a computer program.2 ) = The format for display of pounds/ounces cannot be changed: xx:yy.yyx=lb, y=oz! Some weight units may be blocked from use in legal metrology,depending on national verification laws.44


ISO/GLP-compliant Printout/RecordFeaturesYou can have device information,ID texts and date and time printedbefore (GLP header) and after (GLPfooter) the values of a weighingseries. These parameters include:GLP header:– Date– Time at beginning of measurement– Balance manufacturer– Balance model– Balance serial number– Software version number– Identification number of thecurrent sampling operationGLP footer:– Date– Time at end of measurement– Field for operator signature!Operating the Balance withPrinter:Select the following settings onthe balance and on the printer:– Software handshake:Setup: interf.: handshk.: sftware.(menu code 1. 5. 4. 1)Configuration§ Setting menu codes for the printout(see “Configuration”):– ISO/GLP-compliant printout orrecord only for calibration/adjustment:Setup: Prnt.Out: GLP: Cal.-Adj.(menu code 1. 6. 7. 2) orISO/GLP-compliant printout orrecord always on:Setup: Prnt.Out: GLP: Always on(menu code 1. 6. 7. 3)– Line format for printout: includedata ID codes (22 characters;factory setting):Setup: Prnt.Out: Format: 22 Char.(menu code 1. 6. 6. 2)– Formats for time:setup: prnt.0ut: time: 24 h(menu code 1. 6. 8. 1) orsetup: prnt.oUt: time: 12 h(menu code 1. 6. 8. 2)– Formats for date:setup: prnt.0ut: date: dd.mmm.yy(menu code 1. 6. 9. 1) orsetup: prnt.oUt: date: mmm.dd.yy(menu code 1. 6. 9. 2)!No ISO/GLP-compliant recordis output if any of the followingsettings are configured:Setup: Prnt.Out: Print: Aut. W/o orAut.With (menu code 1. 6. 1. 3, 1. 6.1. 4, ) or Format: 16 Char. (menucode 1. 6. 6. 1)Function KeysTransfer header and first measuredvalue: press PRINT> The header is included with thefirst printout/data record.To output header and referencedata automatically when anapplication program is active:press ENTERExit the application:1) To send the GLP footer:press CLEAR2) Quit application program:press CLEAR again45


The ISO/GLP-compliant printout can contain the following lines:-------------------- Dotted line17-Aug-2005 10:15 Date/time (beginning of measurement)DENVERBalance manufacturerMod. SI-8001 ModelSer. no. 10105355 Balance serial numberVer. no. 00-36-01 Software versionID 2690 923 ID-------------------- Dotted lineL IDMeasurement series no.nRef 10 pcs Counting: reference sample quantitywRef 21.14 g Counting: reference weightQnt + 567 pcs Counting result-------------------- Dotted line17-Aug-2005 10:20 Date/time (end of measurement)Name:Field for operator signatureBlank line-------------------- Dotted lineISO/GLP-compliant printout for external calibration/adjustment:-------------------- Dotted line17-Aug-2005 10:30 Date/time (beginning of measurement)DENVERBalance manufacturerMod. SI-8001 ModelSer. no. 10105352 Balance serial numberVer. no. 00-36-01 Software versionID 2690 923 ID-------------------- Dotted lineCal. Ext. Test Calibration/adjustment modeSet + 5000.0 g Calibration weightDiff. + 0.2 g Difference determined in calibrationCal. Ext. Complete Confirmation of completed calibration procedureDiff. 0.0 g Difference from target following adjustment-------------------- Dotted line17-Aug-2005 10:32 Date/time (end of measurement)Name:Field for operator signatureBlank line-------------------- Dotted line46


Data InterfacePurposeYour balance comes equipped withan interface port for connection to acomputer or other peripheral device.You can use a computer to change,start and/or monitor the functionsof the balance and the applicationprograms.PreparationYou can set these parameters forother devices in the Setup menu(see the chapter entitled “Configuringthe Balance”).You will also find a detailed descriptionof the available data interfacecommands in the file “Data InterfaceDescriptions for S and SI Models”,which you can download fromthe <strong>Denver</strong> website(www.denverinstrument.com).The many and versatile propertiesof these balances can be fully utilizedfor printing out records of the resultswhen you connect your balance to a<strong>Denver</strong> data printer. The recordingcapability for printouts makes it easyfor you to work in compliance withISO/GLP.Hardware HandshakeWith a 4-conductor interface, 1 or 2 characters can be transmitted after CTS.These connections must be made when the balance is connected through theRS-232C port.Pin Assignment Chart for Data Interface<strong>Summit</strong> SeriesStandard RS-232Balance 9-pin port9-pin connectorRxD 2 3 TxDTxD 3 2 RxDDTR 4 4 DTRSignal ground 5 5 Signal groundCTS 8 8 CTS<strong>Summit</strong> SeriesStandard RS-232Balance 9-pin port25-pin connectorRxD 2 2 TxDTxD 3 3 RxDDTR 4 20 DTRSignal ground 5 7 Signal groundCTS 8 5 CTS47


Troubleshooting GuideError codes are shown on the main display for approx. 2 seconds. The program thenreturns automatically to the previous mode.Display Cause SolutionNo segments appear No AC power is available Check the AC power supplyon the displayThe power supply is not Plug in the power supplyplugged inhigh The load exceeds the balance Unload the balancecapacitylow or Err 54 Something is touching Move the object that isthe weighing pantouching the weighing panerr 54, typical Weighing system defect Contact <strong>Denver</strong> Service CenterApp.err. Cannot store data: Increase loadLoad on weighing pan toolight or no sample on panwhile application is activedis.err. Data output not compatible Change the configuration inwith output formatthe operating menuprt.err. Interface port for printer Reset the menu factory settings,output is blockedorContact your local <strong>Denver</strong>Service Centererr 02 Calibration parameter Calibrate only when zeronot met; e.g.:is displayed– balance not tared – Press t to tare the balance– load on weighing pan – Unload the balanceerr 10 The t key is blocked when Press CLEAR to clear the tarethere is data in the second tare memory and release the tare keymemory (net-total); only 1 tarefunction can be used at a timeerr 11 Tare memory not allowed Press tThe weight readout Unstable ambient conditions Set up the balance in anotherchanges constantly (excessive vibration or draft) areaat the place of installationA foreign object is caught Remove the foreign objectbetween weighing pan andbalance housingThe weight readout The balance was not Calibrate/adjust the balanceis obviously wrong calibrated/adjustedBalance not taredTare or zero the balancebefore weighingbefore weighingIf any other errors occur, contact your local <strong>Denver</strong> Service Center.Contact information: http://www.denverinstrument.com48


Care and MaintenanceServiceOn request, <strong>Denver</strong> <strong>Instrument</strong> can provideyou with individual service information.RepairsRepair work must be performed by trainedservice technicians. Any attempt by untrainedpersons to perform repairs may result inconsiderable hazards for the user.Cleaning§ Unplug the AC adapter from the wall outlet(mains supply). If you have an interfacecable connected to the balance port, unplugit from the port.!Make sure that no liquid enters thebalance/scale housing.!Do not aggressive cleaning agents (solventsor similar agents).§ After cleaning, wipe down the balance witha soft, dry cloth.On analytical balances remove and clean theweighing pan as follows:§ Reach beneath the drip/breeeze and liftit carefully, together with the pan support,to avoid damaging the weighing system.!Make sure that no liquid enters the balancehousing.Cleaning Stainless Steel SurfacesClean all stainless steel parts regularly.Remove the stainless steel weighing pan andthoroughly clean it separately. Use a dampcloth or sponge to clean stainless steel partson the balance. You can use any householdcleaning agent that is suitable for use on stainlesssteel. Clean stainless steel surfaces onlyby wiping them down. Then rinse the equipmentthoroughly, making sure to remove allresidues. Afterwards, allow the equipment todry. If desired, you can apply oil to the cleanedsurfaces as additional protection.49


RecyclingSafety InspectionIf there is any indication that safeoperation of the balance is no longerwarranted:§ Turn off the power and disconnectthe equipment from AC powerimmediately.> Lock the equipment in a secure placeto ensure that it cannot be used forthe time being.Notify your nearest <strong>Denver</strong> ServiceCenter. Repair work must be performedby trained service technicians.We recommend having the powersupply inspected by a certified electricianat regular intervals, according to thefollowing checklist:– Insulating resistance: > 7 megaohmsmeasured with a constant voltage ofat least 500 volts at a 500 K-ohm load– Leakage current: < 0.05 mA measuredwith a properly calibrated multimeterInformation and Instructions onDisposal and RepairsPackaging that is no longer requiredmust be disposed of at the local wastedisposal facility. The packaging is madeof environmentally friendly materialsthat can be used as secondary rawmaterials.The equipment, includingaccessories and batteries,does not belong in yourregular household waste.The EU legislation requiresits Member States to collectelectrical and electronic equipmentand disposed of it separately fromother unsorted municipal waste withthe aim of recycling it.In Germany and many other countries,<strong>Denver</strong> <strong>Instrument</strong> takes care of thereturn and legally compliant disposal ofits electrical and electronic equipmenton its own. These products may not beplaced with the household waste orbrought to collection centers run bylocal public disposal operations – noteven by small commercial operators.For disposal in Germany and in theother Member States of the EuropeanEconomic Area (EEA), please contactour service technicians on locationor our Service Center in Goettingen,Germany:<strong>Denver</strong> <strong>Instrument</strong> GmbHRobert-Bosch-Breite 1037079 Goettingen, GermanyIn countries that are not members ofthe European Economic Area (EEA) orwhere no <strong>Denver</strong> affiliates, subsidiaries,dealers or distributors are located,please contact your local authoritiesor a commercial disposal operator.Prior to disposal and/or scrappingof the equipment, any batteries shouldbe removed and disposed of in localcollection boxes.<strong>Denver</strong>, its affiliates, subsidiaries,dealers and distributors will not takeback equipment contaminated withhazardous materials (ABC contamination)– either for repair or disposal.Please refer to the accompanyingleaflet/manual or visit our Internetwebsite (www.denverinstrument.com)for comprehensive information thatincludes our service addresses to contactif you plan to send your equipmentin for repairs or proper disposal.50


OverviewSpecificationsSpecificationsBuilt-in motorizedAll SI modelscalibration weightAC power source/power AC adapter 230 V or 115 25 V,requirements, voltage, frequency+15% to – 20%, 48 – 60 HzPower consumption VA maximum 16; typical 8 (STNG6)Approx. hours of operation withthe SBP01 rechargeable batterypack (backlighting on) h 3551


Specifications for Individual ModelsModel SI-234 | S-234 SI-114 | S-114 SI-64 | S-64Weighing capacity g 230 110 60Readability g 0.0001 0.0001 0.0001Tare range (subtractive) g 230 110 60Repeatability (std. deviation)


Specifications for Individual ModelsModel SI-6002| SI-4002| SI-2002|S-6002 S-4002 S-2002Weighing capacity g 6000 4000 2000Readability g 0.01 0.01 0.01Tare range (subtractive) g 6000 4000 2000Repeatability (std. deviation)


,Verified Models with EC-type Approval: SpecificationsModel SI-234A SI-114A SI-64AType BD ED 100 BD ED 100 BD ED 100Accuracy class 1 ) k k kWeighing capacity, Max 1 ) g 230 110 60Scale interval d 1 ) g 0.0001 0.0001 0.001Tare range (subtractive)< 100% of the maximum capacityVerification scale interval e 1 ) g 0.001 0.001 0.001Minimum capacity, Min 1 ) g 0.01 0.01 0.01Response time (average) s 2.5Range of use according g 0.01 – 230 0.01 – 110 0.01 – 60Allowable ambient °C +17 to +27 (+63°F to +80°F)operating temperatureNet weight, approx. kg 4.8Weighing pan size mm 90 dWeighing chamber height mm 230Dimensions (W + D + H) mm 230 + 303 + 330Model SI-603A SI-403A SI-203AType BD ED 200 BD ED 200 BD ED 200Accuracy class 1 ) K K KWeighing capacity, Max 1 ) g 600 400 200Scale interval d 1 ) g 0.001 0.001 0.001Tare range (subtractive)< 100% of the maximum capacityVerification scale interval e 1 ) g 0.01 0.01 0.01Minimum capacity, Min 1 ) g 0.02 0.02 0.02Response time (average) s 1 1 1Range of use according to CD 1 ) g 0.02 – 600 0.02 – 400 0.02 – 200Allowable ambient °C +10 to +30 (+50°F to +86°F)operating temperatureNet weight, approx. kg 3.6Weighing pan size mm 115 dDimensions (W + D + H) mm 230 + 303 + 1361) CD = Council Directive 90/384/EEC for non-automatic weighing instruments;applicable to the European Economic Area54


Verified Models with EC-type Approval: SpecificationsModel SI-6002A SI-4002A SI-2002AType BD ED 200 BD ED 200 BD ED 200Accuracy class 1 ) K K KWeighing capacity, Max 1 ) g 6000 4000 2000Scale interval d 1 ) g 0.01 0.01 0.01Tare range (subtractive)< 100% of the maximum capacityVerification scale interval e 1 ) g 0.1 0.1 0.1Minimum capacity, Min 1 ) g 0.5 0.5 0.5Response time (average) s 1.1 1.1 1.1Range of use according to CD 1 ) g 0.5 – 6000 0.5 – 4000 0.5 – 2000Allowable ambient °C +10 to +30 (+50°F to +86°F)operating temperatureNet weight, approx. kg 3.5Weighing pan size mm 180 + 180Dimensions (W + D + H) mm 230 + 303 + 91Model SI-602A SI-8001A SI-4001AType BD ED 200 BD ED 200 BD ED 200Accuracy class 1 ) K K KWeighing capacity, Max 1 ) g 600 8000 4000Scale interval d 1 ) g 0.01 0.1 0.1Tare range (subtractive)< 100% of the maximum capacityVerification scale interval e 1 ) g 0.1 1 1Minimum capacity, Min 1 ) g 0.5 5 5Response time (average) s 1.1 1 1Range of use according to CD 1 ) g 0.5 – 600 5 – 8000 5 – 4000Allowable ambient °C +10 to +30 (+50°F to +86°F)operating temperatureNet weight, approx. kg 3.5Weighing pan size mm 180 + 180Dimensions (W + D + H) mm 230 + 303 + 911) CD = Council Directive 90/384/EEC for non-automatic weighing instruments;applicable to the European Economic Area55


AccessoriesDraft ring for round pan unitsIn-use coverPrinter with connecting cableCable with 9-pin connector for balanceWeigh below hanger*Calibration weights: contact your distributor for a complete list* = not allowed in applications involving legal metrology(such as legal for trade)56


C MarkingThe balance complies with the followingEC Directives and EuropeanStandards:Council Directive 89/336/EEC“Electromagnetic compatibility(EMC)”Applicable European Standards:Limitation of emissions:In accordance with product standardEN 61326-1 Class B (residential area)Defined immunity to interference:in accordance with product standardEN 61326-1(minimum test requirements,non-continuous operation)Important Note:The operator shall be responsible forany modifications to <strong>Denver</strong> equipmentand must check and, if necessary,correct these modifications.On request, <strong>Denver</strong> will provide informationon the minimum operatingspecifications (in accordance with theStandards listed above for definedimmunity to interference).73/23/EEC “Electrical equipmentdesigned for use within certainvoltage limits”Applicable European Standards:EN 60950Safety of information technologyequipment including electrical businessequipmentEN 61010Safety requirements for electricalequipment for measurement, controland laboratory use Part 1: GeneralrequirementsIf you use electrical equipmentin installations and under ambientconditions requiring higher safetystandards, you must comply with theprovisions as specified in the applicableregulations for installation in yourcountry.57


InhaltInhalt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60Warn- und Sicherheitshinweise. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61Inbetriebnahme. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62Betrieb . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Anzeige- und Bedienelemente . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Grundfunktion Wägen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69Justieren . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71Voreinstellungen (Menü) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74Funktion der Tasten im Menü. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74Menü-Bedienung; Beispiel: Sprache wählen . . . . . . . . . . . . . . . . . . . . . . . . . . 75Menü: Struktur. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76Parametereinstellungen: Übersicht . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77Eingabe: Ident-Nr., Datum und Uhrzeit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81Anwendungsprogramme . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83Zählen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84Prozentwägen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86Verrechnen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88Tierwägen/Mittelwertbildung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90Netto-Total . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92Summieren . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94Dichtebestimmung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96Einheitenwechsel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98ISO/GLP-Protokoll. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101Datenschnittstelle. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103Fehlermeldungen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104Pflege und Wartung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105Entsorgung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106Übersicht . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107Technische Daten . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107Zubehör . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112C-Kennzeichnung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113Konformitätserklärung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11460


Warn- und SicherheitshinweiseSicherheit§ Diese Betriebsanleitung vor Inbetriebnahmeaufmerksam durchlesen.So können Schäden vermiedenwerden.! Das Gerät nicht in explosionsgefährdetenBereichen einsetzen.! Das Gerät nur von geschultenServicetechnikern öffnen lassen.! Das Gerät vom Netz trennen,bevor Zusatzgeräte angeschlossenoder getrennt werden.! In Betriebsumgebungen mithöheren Sicherheitsanforderungenmüssen die geltenden Errichtungsbestimmungenbeachtetwerden.! Unter extremen elektromagnetischenEinflüssen kann eineBeeinflussung des Anzeigewertesverursacht werden. Nach Endedes Störeinflusses ist das Produktwieder bestimmungsgemäßbenutzbar.Beim Reinigen dürfen keine Flüssigkeitenin die Waage gelangen:nur leicht angefeuchtetes Reinigungstuchverwenden.Installation! Der auf dem Netzgerät aufgedruckteSpannungswert muss mitder lokalen Netzspannung übereinstimmen.– Vorsicht beim Verwenden fremderoder handelsüblicher RS232-Verbindungskabel: Die Pinbelegungensind häufig nicht für<strong>Denver</strong>-Geräte geeignet!Die Belegung sollte anhand derVerbindungspläne überprüft, abweichendeLeitungen solltengetrennt werden.! Weist das Gerät oder das Netzkabelsichtbare Beschädigungenauf, Spannungsversorgungtrennen und Gerät vor weitererBenutzung sichern.– Zubehör von <strong>Denver</strong> ist optimalauf das Gerät angepasst. Daherkeine eigenen Lösungen verwenden.Das Modifizieren des Gerätesund das Anschließen von Fremdkabelnoder -geräten erfolgt aufVerantwortung des Betreibers undist von ihm entsprechend zu prüfen.Hinweise hierzu geben auchunsere Angaben zur Betriebsqualität(gemäß den Normen zurStörfestigkeit), die wir Ihnengerne zur Verfügung stellen.$ Die Waage nicht öffnen. Bei verletzterSicherungsmarke entfälltder Gewährleistungsanspruch.$ Bei Problemen wenden Sie sichbitte an die Kundendienst-Leitstellevon <strong>Denver</strong>.ZeichenerklärungFolgende Symbole werden indieser Anleitung verwendet:§ steht vor Handlungsanweisungen$ steht vor Handlungsanweisungen,die nur unter bestimmtenVoraussetzungen ausgeführtwerden sollen> beschreibt das, was nach einerausgeführten Handlung geschieht– steht vor einem Aufzählungspunkt! weist auf eine Gefahr hin61


InbetriebnahmeLager- und Transportbedingungen– Das Gerät nicht extremen Temperaturen,Stößen, Vibrationen undFeuchtigkeit aussetzen.Auspacken§ Das Gerät nach dem Auspackensofort auf äußere Beschädigungenüberprüfen.$ Im Beschädigungsfall sieheHinweise im Kapitel »Pflege undWartung«, Abschnitt »Sicherheitsüberprüfung«.$ Alle Teile der Originalverpackungfür einen eventuellen Rücktransportaufbewahren. Beim Versand bittekeine Kabel stecken lassen!Lieferumfang– Waage– Waagschale– Unterschale nur bei Modellenmit runder Waagschale– SteckernetzgerätZusätzlich bei S/SI-234, -114, -64:– Schiebetüren-Windschutz– Schirmring– SchirmblechAufstellungBei der Aufstellung Standorte mitden folgenden ungünstigenEinflüssen vermeiden:– Hitze (Heizung, Sonneneinstrahlung)– Direkter Luftzug durch offeneFenster und Türen– Erschütterungen während desWägens– Extreme FeuchtigkeitAkklimatisierenWenn ein kaltes Gerät in eine warmeUmgebung gebracht wird, kann dieszu Betauung (Kondensation) führen.Daher sollte man das vom Netzgetrennte Gerät ca. 2 Stunden beiRaumtemperatur akklimatisieren.Versiegelungsmarke an geeichtenVarianten:Der Gesetzgeber fordert eine Versiegelungder geeichten Waage. DieseVersiegelung erfolgt mittels einerKlebemarke. Wird sie entfernt, erlischtdie Eichgültigkeit und die Waagemuss nachgeeicht werden.Zusätzlich bei S/SI-603, -403, -203:– Glasring-Windschutz(mit Schirmblech und Deckel)Zusätzlich bei SI-203DS:– Schiebetüren-Windschutz62


InstallationWaage aufstellenWaagen mit Schiebetüren-Windschutz§ Teile nacheinander in den Wägeraum setzen:– Schirmblech– Schirmring (nicht bei Modell SI-203DS)– Unterschale– WaagschaleWaagen mit Glasringwindschutz§ Teile nacheinander aufsetzen:– Deckel mit dem Rand nach oben auf dieWaage setzen und drehen, bis er fest sitzt– Unterschale– Waagschale– Glasaufsatz– Deckel mit dem Rand nach untenWaagen mit runder Waagschale§ Teile nacheinander aufsetzen:– Unterschale– WaagschaleWaagen mit eckiger Waagschale§ Waagschale aufsetzen63


Netzanschluss/SchutzmaßnahmenVerwenden Sie nur– Originalnetzgerät für Europa: 6971412§ Winkelstecker an der Waage einsetzen§ Netzgerät an das Stromnetz anschließenNetzgerät mit länderspezifischem Netzkabel:Teilweise werden die Waagen mit einemNetzgerät und länderspezifischen Netzkabelngeliefert.§ Winkelstecker an der Waage einsetzen§ Länderspezifisches Netzkabel auswählenund am Netzgerät montieren§ Netzkabel an das Stromnetz anschließenNetzanschlussleitung an Universalnetzgerätstecken:$ <strong>Denver</strong> Universalnetzgerät mit weitem Netzeingangsspannungsbereich100...240 V~,Best.-Nr. 6971966 und austauschbarerNetzanschlussleitung:6900900 (Euro) 6900905 (AUS)6900901 (US/CDN) 6900902 (ZA)6971945 (UK) 6971776 (Italien)6971980 (Dänemark) 6971975 (Israel)64SchutzmaßnahmenBei Steckernetzgerät 6971412:Das Netzgerät der Schutzklasse 2 darf ohneweitere Maßnahmen an jede Steckdoseangeschlossen werden.Bei Universalnetzgerät 6971966:Das Netzgerät der Schutzklasse 1 darf an jedeSteckdose mit Schutzleiteranschluss (PE)angeschlossen werden.Die Ausgangsspannung ist durch einen Pol mitdem Waagengehäuse verbunden. Das Waagengehäusedarf zu Funktionszwecken geerdetwerden. Die Datenschnittstelle ist ebenfalls galvanischmit dem Waagengehäuse (Masse) verbunden.


Anschluss von elektronischen Komponenten(Peripherie)§ Vor Anschluss oder Trennen von Zusatzgeräten(Drucker, PC) an die Datenschnittstellemuss die Waage vom Netz getrennt werden.AnwärmzeitUm genaue Resultate zu liefern, braucht dasGerät eine Anwärmzeit von 30 Minuten. Erstdann ist die notwendige Betriebstemperaturerreicht.Geeichte Waagen im eichpflichtigen Verkehreinsetzen:$ Anwärmzeit von mindestens 24 Stundeneinhalten nach erstmaligem Anschluss andas Stromnetz.Betrieb außerhalb des Temperaturbereichs!Einsatz der Waage außerhalb des Temperaturbereichsvon +10…+30°C (50°…86°F):Abweichungen von den im Kapitel »TechnischeDaten« aufgeführten Spezifikationensind möglich.65


UnterflurwägungFür Wägungen unterhalb der Waage steht eineUnterflurwägeeinrichtung zur Verfügung.$ Für eichpflichtige Anwendung nicht zulässig.§ Verschlussplatte am Waagenboden herausheben.Achtung: Die Waage dabei nur auf die Seiteund nicht auf den Kopf legen!§ Fester Haken 1: Wägegut (z.B. mit einemDraht) an dem Haken einhängen.21$ Bohrung 2 (nicht bei Modellen: S/SI-203, -602, -402, -8001, -4001): Speziellen Hakenvorsichtig einschrauben oder einen Hakendirekt bei <strong>Denver</strong> anfordern.$ Ggf. Abschirmung gegen Luftzug installieren.66


Waage nivellierenZweck:– Ausgleich von Unebenheiten des Waagen-Stellplatzes.Die Waage nach jedem Stellplatzwechsel neunivellieren. Das Nivellieren erfolgt nur mit denbeiden vorderen Stellfüßen.§ Beide hintere Stellfüße eindrehen (nur beiModellen mit rechteckiger Waagschale).§ Vordere Fußschrauben gemäß Abbildungso drehen, bis die Luftblase der Libelle inder Kreismitte steht.> In der Regel sind mehrere Nivellierschrittenötig.§ Bei Modellen mit rechteckiger Waagschale:Beide hintere Stellfüße herausdrehen, bissie die Aufstellfläche berühren.DiebstahlsicherungZur Diebstahlsicherung die Befestigungsösean der Rückseite der Waage verwenden.§ Waage am Aufstellort z.B. mit einer Ketteoder einem Schloss befestigen.67


BetriebAnzeigeelemente im Überblick121211Pos.10987Bezeichnung1 Wägeeinheiten2 Anzeige der Menüebene3 Piktogramm für »GLP-Ausdruck aktiv«4 Piktogramm für »Drucken aktiv«5 Anwendungsprogramm aktiv6 Kennzeichnung: Kein Wägewert7 Anzeige: Brutto- und Nettowert8 Piktogramme für eingestellte Anwendung(W, Z, L, V, R, A, C)9 Anzeige: Justierfunktion10 Piktogramm für Nullbereich(nur bei geeichten Modellen)Bedienelemente im Überblick3456Pos. Bezeichnung11 Busy-Symbol: Kommandoverarbeitungaktiv (z.B. »Warten auf Stillstand«);Nach dem Einschalten wird A bis zumersten Tastendruck angezeigt.12 Gewichtswertanzeige entsprechendgewählter BasiseinheitSymbol: Nächsten Unterpunkt innerhalb einerMenüebene wählen↵ Menüpunkt bestätigen87165Pos.4Bezeichnung1 Tarieren2 Anwendungsprogramm starten3 Umschalten im Anwendungsprogramm |Menü aufrufen4 Löschen (Clear Function)Diese Taste wird allgemein als Abbruchtastebenutzt:– Anwendungsprogramme beenden– Gestartete Justiervorgänge abbrechen |Menü verlassen6823Pos. Bezeichnung5 Justiervorgänge starten6 Datenausgabe:Diese Taste aktiviert die Ausgabe derAnzeigewerte über die integrierteDatenschnittstelle.7 Ein-/Ausschalten8 Libelle


Grundfunktion WägenMerkmale– Waage tarieren– Wägewert druckenVorbereitung§ Waage einschalten:Taste drücken§ Ggf. Waage tarieren:Taste t drücken$ Ggf. Voreinstellungen ändern:siehe Kapitel »Voreinstellungen«$ Ggf. Werksvoreinstellungen laden:siehe Kapitel »Voreinstellungen«Weitere Funktion:$ Waage ausschalten:Taste drücken$ Waage im Standby-Betrieb:Anzeige der aktuellen UhrzeitEinsetzen der geeichten Waage imeichpflichtigen Verkehr:Die Bauartzulassung zur Eichung giltnur für nichtselbsttätige Waagen; fürselbsttätigen Betrieb mit oder ohnezusätzlich angebaute Einrichtungensind die für den Aufstellort geltendennationalen Vorschriften zu beachten.§ Vor dem Einsatz im eichpflichtigenVerkehr die Waage am Aufstellortmit Hilfe der eingebauten Justiergewichtsschaltungjustieren: sieheAbschnitt »Kalibrieren, Justieren«in diesem Kapitel.$ Auf dem Kennzeichnungsschildangegebener Temperaturbereich(°C) darf beim Betrieb nicht überschrittenwerden.Beispiel:BD ED 200K +10…+30°C69


BeispielWägewert ermittelnSchritt Taste drücken Anzeige/AusgabeWaage im Standby-Betrieb11:54 AM1. Waage einschalten 0.0 gSelbsttest wird durchgeführt.Danach führt die Waage eineEinschalttarierung durch.2. Behälter für Wägegut auflegen + 11.5 g(hier z.B. 11,5 g).3. Waage tarieren t 0.0 g4. Wägegut in Behälter legen + 132.0 g(hier z.B. 132 g)Das Gewichtseinheiten-Symbol»g« erscheint, wenn dasErgebnis stabil ist5. Wägewert drucken PRINTN + 132.0 g70


JustierenZweckJustieren ist die Tätigkeit, um dieAbweichung zwischen dem angezeigtenMesswert und dem wahrenMassewert zu beseitigen, bzw. aufdie zulässigen Fehlergrenzen zureduzieren.Einsatz der geeichten Waage im eichpflichtigenVerkehr:Vor dem Einsatz im eichpflichtigenVerkehr die Funktion »Intern Justieren«am Aufstellort durchführen.Extern Justieren bei geeichtenWaagenIm eichpflichtigen Verkehr ist extern-Justieren gesperrt:– Schaltereinstellung verriegelt– Schalterabdeckung versiegeltDie Waage kann für Justierengesperrt werden:– Cal.Just. - gesprt. wählen– Eichschalter an der Rückseite derWaage schließenMerkmaleDer Justiervorgang kann nur gestartetwerden, wenn– die Waage unbelastet ist– die Waage tariert ist– das interne Wägesignal stabil ist– der Wert des aufgelegten Gewichtesdarf nur um maximal 2% vomSollwert abweichen.Sind diese Voraussetzungen nichterfüllt, erscheint eine Fehlermeldung»Err 02«.Das Justieren kann mit unterschiedlichenGewichtseinheiten erfolgen:CAL.Einh. - gramm, Kilogr. oder Pound(entfällt bei geeichten Waagen)Die Ergebnisse vom Justierenkönnen in einem ISO/GLP-Protokollausgedruckt werden, siehe Seite 95.Nach dem Justieren ist das Anwendungsprogrammgelöscht.Intern JustierenIm Menü muss der Punkt Cal.Just. -cal.int. eingestellt sein.Im Waagengehäuse befindet sich einJustiergewicht, das motorisch internaufgelegt werden kann.§ Justieren wählen: Taste q> Internes Justiergewicht wird automatischaufgelegt> Waage wird justiert> Waage wird entlastet vom internenGewicht71


Intern Justieren (nur bei SI-Modellen)Voreinstellung:SETUP - Waage - CAL.Just. - Cal.Int. (Code 1.1. 9. 4)Im Waagengehäuse befindet sich ein Justiergewicht, das motorisch intern aufgelegtwerden kann.Schritt Taste drücken Anzeige1. Waage tarieren t 0.0 g2. Justiervorgang starten q CAL.Int.Internes Gewicht wirdautomatisch aufgelegt.CAL.RUN.3. Justieren ausgeführt CAL.end4. Waage wird entlastet vominternen Gewicht0.0 g72


Extern JustierenVoreinstellung:SETUP - Waage - CAL.Just. - Cal.Ext. (Code 1.1. 9. 1)Werkseitig ist ein Gewichtswert voreingestellt (siehe Abschnitt »TechnischeDaten«)Schritt Taste drücken Anzeige1. Waage tarieren t 0.0 g2. Justiervorgang starten q Cal.Ext.Nach Übernahme des Nullpunktswird das aufzulegende Justiergewichtblinkend angezeigt.- 5000.0 g3. Angezeigtes Justiergewicht 5000.0 gauflegen (hier z.B. 5000 g).Gewicht zu klein:Vorzeichen »–« erscheintGewicht zu groß:Vorzeichen »+« erscheintBlinken endet, wenn das Justiergewichtinnerhalb der vorgegebenenGrenzen liegt.4. Justieren ausgeführt; Cal.enddanach erscheint dasJustiergewicht+ 5000.0 g5. Justiergewicht abnehmen 0.0 g73


Voreinstellungen (Menü)Konfiguration der Waage, d.h. Anpassung an die Anforderungen der Benutzer.Funktion der Tasten im Menü:Anzeigesymbol Taste FunktionVSELECTMenüpunkt einstellenlanger Tastendruck> ENTERNächsten Unterpunkt einerMenüebene wählen(mit Cursor nach rechts bis zu4 Menüebenen)↵ENTERCLEARMenüpunkt bestätigenMenü verlassen und Einstellunglanger Tastendruck speichern von jeder Position


Menü-BedienungBeispiel: Sprache einstellenSchritt Taste drücken Anzeige1. Einstieg ins Menü: SELECT lang Applic.Im Wägemodus den1. Menüpunkt anzeigen2. Aufwärts umlaufend innerhalb mehrmals SELECT Inputeiner Menüebene bewegen; ...nach dem letzten Menüpunktlanguag.erscheint wieder der ersteMenüpunkt (Scrollen)3. Nächste Menüebenen wählen mehrmals ENTER english o(nach rechts umlaufend)4. Einstellung ändern: SELECTDeutschAufwärts umlaufend denMenüpunkt wählen5. Einstellung bestätigen; ENTERDeutsch o»o« zeigt den eingestelltenMenüpunkt an6. Zurück zur vorherigen Menü- CLEARSpracheebene (von der 2. Menüebene)$ Ggf. weitere Menüpunkteeinstellen SELECT , ENTER7. Einstellung speichern mehrmalsund Menü verlassenCLEARoder$ Parametereinstellung ohneAbspeichern verlassen> Neustart der Anwendung 0.0 g75


Menüstruktur (Übersicht)Ebene 1 Ebene 2 Ebene 3 CodesSetup waage Umgeb.ungsbedingungen 1. 1. 1.Waagenparameter app.fil. Anwendungsfilter 1. 1. 2.stillst.andsbereich 1. 1. 3.Tara Tarierung 1 ) 1. 1. 5Aut.zero Autozero 1. 1. 6Einheit Basis-Gewichtseinheit 1. 1. 7.stellen Anzeigegenauigkeit 1 ) 1. 1. 8.Cal./jus. Funktion der Taste q 1. 1. 9.cal.einh.eit für Justiergewicht 1 ) 1. 1.11.Interf. Schnittstelle Baudrate 1. 5. 1.parity Parität 1. 5. 2.Stopbit Anzahl Stopbits 1. 5. 3.handshk. Handshake Betriebsart 1. 5. 4.databit Anzahl Datenbits 1. 5. 5.dat.prot. SBI (ASCII) oder Drucker 1. 5. 6.protok.oll (Drucken) druck (manuell/automatisch) 1. 6. 1.abbruch des autom. Druckens 1. 6. 2.tar./prt. Waage tarieren nach Einzelprint 1. 6. 4.init.Dat. Ausdruck der Anwendungsparam. 1. 6. 5.format Zeilenformat des Ausdr. 1. 6. 6.GLP Ausdruck als ISO/GLP-Protokoll 1. 6. 7.uhzeit: 12h/24h 1. 6. 8.datum-Format 1. 6. 9.extras menue Menü nur lesbar/änderbar 1. 8. 1.(Zusatzfunktionen) Hupe Akustisches Signal 1. 8. 2.tasten (Tastatur) 1. 8. 3.on-mode Einschalten der Waage 1. 8. 5.b.leucht. Hinterleuchtung der Anzeige 1. 8. 6.reset menue Werkseinstellung 1. 9. 1.Anwend.ungs- wiegen 2. 1.programme einheitenwechsel Stellen Anzeigegenauigkeit 1 ) 2. 2. 2.Zaehlen aufloes.ung 2. 3. 1.optim. Autom. Referenzoptimierung 2. 3. 2.prozentwägen N.komma Nachkommastellen 2. 4. 1.net.tot. Netto-Total Ko.Druck Komponentendruck 2. 5. 1.Summe Summieren Ko.Druck Komponentendruck 2. 6. 1.tierwg. Tierwägen Tierbew. Tierbewegung 2. 7. 1.start 2. 7. 2.Verrech. Verrechnen Methode (Operator) 2. 8. 1.N.komma Nachkommastellen 2. 8. 2.Dichtebestimmung N.komma Nachkommastellen 2. 9. 1.Input Eingabe ident.Nr., datum, uhrzeit Eingabe: ID-Nr., Datum, Uhrzeit 3. 1./2./3.InFormation Ver. nr. , Ser. nr. , modell Anzeige der Softw.-Vers., Serien-Nr., Modell 4. 1./2./3.Sprache English (Werksvoreinstellung) 5. 1.(Languag.) Deutsch 5. 2.franc.çais 5. 3.ital.iano 5. 4.Espanol 5. 5.codes Darstellung des Menüs als Codes 5. 6.1 ) Einstellungsänderung entfällt bei geeichten Waagen76


Parametereinstellungen: Übersichtο = Werksvoreinstellung √ = Einstellung BenutzerEbene 1 Ebene 2 Ebene 3 Ebene 4 CodeSetup waage Umgeb.ungs- Sehr .Ruh.ig 1. 1. 1. 1Waagen- bedingungen ο Ruhig 1. 1. 1. 2parameter (Filteran- unRuhig 1. 1. 1. 3passung) Sehr.unr.uhig 1. 1. 1. 4app.fil. An- ο Auswaeg.en 1. 1. 2. 1wendungsfilter Dosier.en 1. 1. 2. 2stillst.ands- 1/4-digi.t (Ziffernschritt) 1. 1. 3. 1bereich 1/2-digi.t (Ziffernschritt) 1. 1. 3. 21-digi.t (Ziffernschritt) 1. 1. 3. 3ο 2-digi.t (Ziffernschritte) 1. 1. 3. 44-digi.t (Ziffernschritte) 1. 1. 3. 58-digi.t (Ziffernschritte) 1 ) 1. 1. 3. 6Tara 1 ) Ohne.std. (Stillstand) 1. 1. 5. 1Tarierung ο nach.std. (Stillstand) 1. 1. 5. 2Aut.zero Aus 1. 1. 6. 1Autozero ο ein 1. 1. 6. 2Einheit Einheiten siehe Kapitel 1. 1. 7. 1Basis- »Einheitenwechsel« bisGewichtseinheit 1. 1. 7.23stellen 1 ) ο alle 1. 1. 8. 1Anzeige- Minus 1 1. 1. 8. 2genauigkeitCal./jus. ο cal.ext. Extern Justieren 1 ) 1. 1. 9. 1Funktion der cal.int. Intern Justieren 2 ) 1. 1. 9. 2Taste q gesprt. q gesperrt 1. 1. 3. 3cal.einh.eit 1 ) ο Gramm 1. 1.11. 1für Justier- Kilogr.amm 1. 1.11. 2gewicht pound 1. 1.11. 31 ) Einstellungsänderung entfällt bei geeichten Waagen2 ) nur bei SI-Modellen77


Ebene 1 Ebene 2 Ebene 3 Ebene 4 CodeSetup Interf. Baudrate 600 1. 5. 1. 3Schnittstelle ο 1200 1. 5. 1. 42400 1. 5. 1. 54800 1. 5. 1. 69600 1. 5. 1. 719200 1. 5. 1. 8Parity ο Odd 1. 5. 2. 3Parität Even 1. 5. 2. 4None 1. 5. 2. 5StopBit ο 1 Stop 1. 5. 3. 1Anzahl Stopbits 2 Stop 1. 5. 3. 2Handshk. Softw.are 1. 5. 4. 1Handshake ο Hardw.are 1. 5. 4. 2Betriebsart None 1. 5. 4. 3DataBit ο 7 Bits 1. 5. 5. 1Anzahl Datenbits 8 Bits 1. 5. 5. 2Dat.Rec. Kom- SBI (ASCII) 1 ) 1. 5. 6. 1munikationsart ο drucker (GLP-Protokoll) 1. 5. 6. 2protok.oll druck Man.ohne Stillstand 1. 6. 1. 1(Drucken) (manuell/ ο man.mit. Stillstand 1. 6. 1. 2automatisch) auto.ohn.e Stillstand 1. 6. 1. 3auto.mit. Stillstand 1. 6. 1. 4abbruch des ο aus Abbruch nicht möglich 1. 6. 2. 1autom. Druckens ein Abbruch mit Taste PRINT 1. 6. 2. 2tar./prt. ο aus 1. 6. 4. 1Waage tarieren ein 1. 6. 4. 2nach Einzelprint1 ) Hinweis zu geeichten Waagen:In der Einstellung »SBI« erfolgt keine automatische Kennzeichnung der nichtgeeichten Anzeigestelle.Entsprechende Maßnahmen oder Einstellungen dafür an der Zusatzeinrichtung durchführen.78


Ebene 1 Ebene 2 Ebene3 Ebene 4 CodeSetup protok.oll init.Dat. Ausdr. aus 1. 6. 5. 1(Drucken) der Anwen- ο alle Parameter 1. 6. 5. 2dungsparameter hauptp.arameter 1. 6. 5. 2format Zeilen- 16. zei.chen (ohne Kennz.) 1. 6. 6. 1format des Ausdr. ο 22. Zei.chen (mit Kennzeich.) 1. 6. 6. 22 Zeil.en (Datum/Uhrzeit 1. 6. 6. 3und Wägewert)GLP Ausdruck ο aus 1. 6. 7. 1als ISO/GLP- cal./Jus. nur bei Justieren 1. 6. 7. 2Protokoll Immer an 1. 6. 7. 3uhrzeit ο 24h-Anzeige 1. 6. 8. 112h-Anzeige »AM/PM« 1. 6. 8. 2Datum ο tt.mmm.jj-Format 1. 6. 9. 1mmm.tt.jj-Format 1. 6. 9. 2extras menue aenderb.ar 1. 8. 1. 1(Zusatz- lesbar Parameter nur lesen 1. 8. 1. 2funktionen)Hupe Akus- aus 1. 8. 2. 1tisches Signal ο ein 1. 8. 2. 2tasten ο frei 1. 8. 3. 1(Tastatur) gesprt. 1. 8. 3. 2on-mode ο off/on Aus/ein/Stand-by 1. 8. 5. 1Einschalten standby Ein/Stand-by 1. 8. 5. 2der Waage Auto-On Automatisch ein 1. 8. 5. 3b.leucht. aus 1. 8. 6. 1Hinterleuchtung ο ein 1. 8. 6. 2der Anz.reset menue ja wiederherstellen 1. 9. 1. 1Menü-Reset Werkseinstellung ο Nein nicht wiederherstellen 1. 9. 1. 279


Ebene 1 Ebene 2 Ebene3 Ebene 4 Codeanwend. wiegen 2. 1.-ungspro- einheiten- stellen 1 ) ο alle 2. 2. 2. 1gramme wechsel Anzeige- Minus 1 2. 2. 2. 2genauigkeitZaehlen aufloes.ung ο anz.gen. Anzeigegenau 2. 3. 1. 110 fach genauer als Anz. 2. 3. 1. 2optim. Auto. ο aus 2. 3. 2. 1Referenz- autom.atisch 2. 3. 2. 2optimierungprozent N.komma ohne Nachkommastelle 2. 4. 1. 1Prozent- Nachkomma- ο 1 nks. 1 Nachkommastelle 2. 4. 1. 2wägen stellen 2 nks. 2 Nachkommastellen 2. 4. 1. 33 nks. 3 Nachkommastellen 2. 4. 1. 4net.tot. Ko.Druck aus 2. 5. 1. 1Netto-Total Komponenten- ο Ein 2. 5. 1. 2druckSumme Ko.Druck aus 2. 6. 1. 1Summieren Komponenten- ο Ein 2. 6. 1. 2drucktierwg. Tierbew. wenig (ruhig, 2% v. Messobj.) 2. 7. 1. 1Tierwägen Tierbewegung ο mittel (normal, 5% v. Mess.) 2. 7. 1. 2viel (unruhig, 20% v. Mess.) 2. 7. 1. 3start manuell 2. 7. 2. 1ο autom.atisch 2. 7. 2. 2Verrech. Methode ο Multi.plikator 2. 8. 1. 1Verrechnen (Operator) divis.or 2. 8. 1. 2N.komma ohne Nachkommastelle 2. 8. 2. 1Nachkomma- ο 1 nks. 1 Nachkommastelle 2. 8. 2. 2stellen 2 nks. 2 Nachkommastellen 2. 8. 2. 33 nks. 3 Nachkommastellen 2. 8. 2. 4Dichte N.komma ohne Nachkommastelle 2. 9. 1. 1-bestimmung Nachkomma- ο 1 nks. 1 Nachkommastelle 2. 9. 1. 2stellen1 ) Einstellungsänderung entfällt bei geeichten Waagen80


Eingabe: Ident-Nummer, Datum und UhrzeitEbene 1 Ebene 2 Ebene 3 CodeInput ident.Nr. ID-Eingabe max. 7-stellig 3. 1.EingabeMögliche Zeichen: 0-9, A-Z,Strich- und LeerzeichenDATum Datum einstellen 3. 2.Uhrzeit Uhrzeit einstellen 3. 3.Je nach Einstellung des Menüpunktes »SETUP – PROTOK.oll – DATUM«erfolgt die Anzeige des Datums im folgenden Format:FormatDD.MMM.YYAnzeige: DatumMMM.DD.YYJe nach Einstellung des Menüpunktes »SETUP – PROTOK.oll – UHRZEIT«erfolgt die Anzeige der Uhrzeit im folgenden Format:Zeit24h-ModeAnzeige: Uhrzeit12h-Mode81


Beispiel: Ident-Nr., Datum und Uhrzeit einstellenSchritt Taste drücken Anzeige1. Einstieg ins Menü: SELECT lang Anwend.1. Menüebene anzeigen2. »Input« wählen SELECTInput3. Eingabe für Ident-Nr. wählen 2x ENTERident.nR.4. Ident-Nr. einstellen oder ändern – SELECT3------mit autom. Tastenwiederholung: SELECT lang5. Innerhalb der 7-stelligen ENTER oder 3-aBC12Ident-Nr. bewegenCLEAR6. Eingabe speichern an der ENTERident.nR.letzten Position der Ident-Nr.7. »Datum« wählen und bestätigen SELECT ENTER, 08.feb.068. Einstellung ändern – SELECT10.feb.06mit autom. Tastenwiederholung: SELECT lang9. Position wechseln zwischen ENTER oder 10.feb.06Tag/Monat/JahrCLEAR10. Einstellung speichern auf derPosition »Jahr«datum11. »Uhrzeit« wählen SELECT ENTER, 10.46.2312. Einstellung ändern – SELECT11.46.23mit autom. Tastenwiederholung SELECT lang13. Position wechseln zwischen ENTER oder 11.46.32Stunde/Minute/SekundeCLEAR14. Sekunde auf »Null« setzen SELECT11.47.0015. Einstellung speichern auf derPosition »Sekunde«16. Einstellung speichern mehrmalsund Menü verlassenCLEARuhrzeit> Neustart der Anwendung 0.0 gENTERENTER82


Gerätespezifische InformationenEbene 1 Ebene 2 Ebene 3 Beispiel CodeInFo Ver. nr. Anzeige der Software-Version rel.36.01 4. 1.rmationSer. nr. Anzeige der Seriennummer 10801234 4. 2.(zwischen Anzeige vomoberen Teil/unteren Teilwechseln:Taste SELECT drücken)modell Anzeige der Modell SI-8001 4. 3.bezeichnung (zwischenAnzeige vom oberen, mittlerenund unteren Teil wechseln:Taste SELECT drücken)Menü-Darstellung: Sprache oder Codes wählenSprache English (Werksvoreinstellung) 5. 1.(Languag.) Deutsch 5. 2.franc.çais 5. 3.ital.iano 5. 4.Espanol 5. 5.codes Darstellung des Menüs als Codes 5. 6.AnwendungsprogrammeHinweis zu geeichten Waagen:Alle Anwendungsprogramme könnenbei einer geeichten Waage angewähltwerden.Errechnete Werte können mitfolgenden Zeichen gekennzeichnetwerden:– Prozent = %– Stückzahl (Zählen) = pcs– Rechenwerte = o, G83


ZählenAnzeigensymbol: ZZweckBeim Zählen kann die Anzahl vonTeilen ermittelt werden, die ein annäherndgleiches Stückgewicht haben.Dabei wird zunächst das Gewichteiner manuell bestimmten Anzahlvon Teilen bestimmt und daraus dasEinzelgewicht errechnet (Referenz).Aus dem Gewicht der unbekanntenZahl von Teilen wird dann dieGesamtzahl berechnet.Vorbereitung§ Anwendungsprogramm »Zählen«im Menü einstellen:siehe Kapitel »Voreinstellungen«§ Parameter einstellen:Anwend.ungsprogrammeZaehlenAufloes.ungo Anz.Gen. Anzeigegenau10-fach 10-fach genaueroptim. Autom. Referenzoptimierungo Ausautom.o = WerkseinstellungAnzeigegenauAutomatischReferenzstückzahl ändernFunktion aufrufen:Taste SELECT drückenMögliche Referenzstückzahl 1 bis 100wählen:In Einer-Schritten: Taste SELECT kurzdrückenIn 10er-Schritten:Taste SELECT gedrückt halten.Die gewählte Stückzahl wirdnetzausfallsicher gespeichert.Optimierung der ZählergebnisseDie automatische Referenzoptimierungführt beim Zählen zu genauerenResultaten. Diese Funktion kann imMenü ein- oder ausgeschaltetwerden.Eine automatische Referenzoptimierungwird durchgeführt, wenn dieVoraussetzungen und das eingestellteStillstandskriterium erfüllt sind.Mit der Optimierung opt wird kurzzeitigdas neue mittlere Stückgewichtangezeigt.Ausdruck ZählennRef 10 : ReferenzstückzahlwRef 21.14 g : Referenzgewichtfür 1 StückQnt + 500 pcs : Errechnete Stückzahl84


Beispiel: Zählen von gleichgewichtigen TeilenVoreinstellungen der Parameter: Anwend. - Zaehlen (Code 2. 3.)Schritt Taste drücken Anzeige/Ausgabe1. Behälter zum Einfüllen der Teile + 22.6 gauf die Waage stellen2. Waage tarieren t 0.0 g3. Referenzstückzahl in denBehälter legen (hier: 20 Stück)4. Referenzstückzahl ändern SELECTref 10 pcs5. Referenzstückzahl wählen: mehrmalsIn Einer-Schritten (1, 2, 3, ..., 100)SELECT kurzIn 10er-Schritten (10, 20, ..., 100) oder SELECTref 20 pcsgedrückt halten6. Gewählte Referenzstückzahl ENTER+ 20 pcsbestätigen und Anwendung nRef*20 pcsstarten. Das aktuelle Referenz- wRef 1.07 ggewicht bleibt gespeichert, bisdie Referenz neu gesetzt oderdie Stromversorgung unterbrochenwird.7. Gewünschte Teilemenge + 500 pcseinfüllen8. Ggf. Stückzahl drucken PRINTQnt + 500 pcs9. Anzeige umschalten zwischen + 1.07 g Gmittlerem Stückgewicht, mehrmals + 535.0 g *Gewicht, Stückzahl SELECT+ 500 * pcs*10. Waage entlasten – 21 pcs *11. Ggf. weiter bei 7.12. Referenzwert löschen CLEAR0.0 g85


ProzentwägenAnzeigensymbol: %ZweckMit diesem Anwendungsprogrammkann der prozentuale Anteil einesWägegutes bezogen auf ein Referenzgewichtermittelt werden.Vorbereitung§ Anwendungsprogramm »Prozentwägen«im Menü einstellen:siehe Kapitel »Voreinstellungen«Referenzprozentwert ändernFunktion aufrufen:Taste SELECT drückenMögliche Referenz 1 bis 100 wählen:In Einer-Schritten: Taste SELECT kurzdrücken10er-Schritten: Taste SELECT gedrückthalten.Die gewählte Prozentzahl wird netzausfallsichergespeichert.§ Parameter einstellen:Anwend.ungsprogrammeproZent Prozentwägenn.komma. Nachkommastelleno = Werkseinstellungohne Nachkommastelleo 1nks. 1 Nachkommastelle2nks. 2 Nachkommastellen3nks. 3 NachkommastellenAusdruck ProzentwägenpRef 100 : ReferenzprozentzahlWxx% 111.6 g : Referenzgewichtnetto xx% fürgewählte ReferenzprozentzahlPrc + 94.9 % : ErrechneteProzentzahl86


Beispiel: Restgewichtsbestimmung in ProzentVoreinstellung der Parameter: Anwend. - prozent (Code 2. 4.)Referenzprozentwert: Ref 100%Schritt Taste drücken Anzeige/Ausgabe1. Waage tarieren t 0.0 g2. Vorbereitete Probe für 100%auflegen (hier: 111,6 g)3. Information:Referenzprozentzahl eingeben SELECTref 100 %(Ändern: siehe vorherige Seite)4. Waage initialisieren. ENTER+ 100.0 %Das aktuelle Referenzgewicht bleibt pRef*100 %gespeichert, bis die Referenz neu Wxx% + 111.6 ggesetzt oder die Stromversorgungunterbrochen wird.5. Probe abnehmen(z.B. Probe trocken)6. Unbekanntes Gewicht auflegen + 94.9 % *(hier 105,9 g)7. Ggf. Prozentwert drucken PRINTPrc + 94.9 %8. Anzeige umschalten zwischen mehrmals + 105.9 gGewicht und Prozentzahl SELECT+ 94.9 * %*9. Restgewichtsanzeige undReferenzprozentzahl löschen CLEAR+ 105.9 g10. Ggf. Netto-Restgewicht drucken PRINTN + 105.9 g87


VerrechnenAnzeigensymbol: CZweckMit diesem Anwendungsprogrammkann der Wägewert mit Hilfe einesMultiplikators oder Divisors verrechnetwerden.Einsatzgebiete sind zum Beispiel FlächengewichtsbestimmungenvonPapier.Vorbereitung§ Anwendungsprogramm»Verrechnen« im Menü einstellen:siehe Kapitel »Voreinstellungen«Faktor oder Divisor einstellenFunktion aufrufen:Taste SELECT drückenMöglicher Zahlenumfang, siebennumerische Stellen und ein Dezimalpunktwählen (0.000001 – 9999999):In Einer-Schritten:Taste SELECT kurz drückenoder umlaufend:Taste SELECT gedrückt halten.Der gewählte Operator wird netzausfallsichergespeichert.§ Parameter einstellen:Anwend.ungsprogrammeverrech. Verrechnenmethode (Operator)o Multi. Multiplikatordivis. Divisorn.komma. Nachkommastellenohne Nachkommastelleo 1nks. 1 Nachkommastelle2nks. 2 Nachkommastellen3nks. 3 NachkommastellenAusdruck VerrechnenMul + 1.2634 : MultiplikatorDiv + 0.6237 : DivisorRes + 79.7 o : Ergebniso = Werkseinstellung88


Beispiel:Berechnung des Flächengewichtes von Papier: Das Flächengewicht einesDIN A4 Blattes mit der Fläche 0,210 m + 0,297 m = 0,06237 m 2 soll ermitteltwerden. Das Flächengewicht ergibt sich aus der Division des Gewichtswertesdurch die Fläche.Voreinstellungen für Methode Divisor:Anwend. - verrech. - methode - Divis. (Code 2. 8. 1. 2)Schritt Taste drücken Anzeige/Ausgabe1. Waage tarieren t 0.00 g2. Eingabe Divisor wählen SELECT______0.3. Divisor einstellen (hier: 0,06237): ENTER ,Dezimalpunkt positionieren, 5+ SELECT , __.000002+ ENTER ,Ziffern in Einer-Schritten oder mehrmals __.06000umlaufendSELECT kurz odergedrückt halten,ENTER , usw. __.062374. Einstellung Divisor übernehmenund Waage initialisieren. ENTER+ 0.0 oDer aktuelle Divisor bleibt Div 0.6237netzausfallsicher gespeichert,bis die Einstellung geändert wird.5. Flächengewicht ermitteln: + 79.7 o *DIN A4 Blatt auflegen6. Ggf. Flächengewicht drucken PRINTRes + 79.7 o7. Anzeige umschalten zwischen mehrmals + 4.97 gGewicht und verrechnetem Wert SELECT+ 79.7 * o*8. Waage entlasten + 0.0 o *9. Ggf. weiter bei 5.89


Tierwägen/MittelwertbildungAnzeigensymbol: VZweckMit diesem Anwendungsprogrammist die Messung unruhiger Wägegüter(z.B. Tiere) oder die Messung vonWägegütern in extrem unruhigerUmgebung möglich. Dafür wird übermehrere Messzyklen ein Mittelwertgebildet.Vorbereitung§ Anwendungsprogramm»Tierwägen« im Menü einstellen:siehe Kapitel »Voreinstellungen«Anzahl der Untermessungen ändernFunktion aufrufen:Taste SELECT drückenMögliche Messungen 1 bis 100wählen:In Einer-Schritten: Taste SELECT kurzdrücken10er-Schritten: Taste SELECT gedrückthalten.Die gewählte Anzahl der Untermessungenwird netzausfallsichergespeichert.§ Parameter einstellen:Anwend.ungsprogrammetierwaeg. Tierwägentierbew. TierbewegungStartwenig (ruhig)o mittel (normal)viel (unruhig)manuello Autom.atischAusdruck TierwägenmDef 20 : Definierte Anzahlder Untermessungenx-Net + 410.1 g : Ergebnis der Mittelwertbildungo = Werkseinstellung90


Beispiel: Automatisches Wägen von Tieren mit 20 UntermessungenVoreinstellungen für Parameter: Anwend. - Tierwaeg. (Code 2. 7.)Schritt Taste drücken Anzeige/Ausgabe1. Tierwaagschale aufstellen 22.6 g2. Waage tarieren t 0.0 g3. Anzahl der Untermessungenändern SELECTref 304. Messungen wählen: mehrmalsIn Einer-Schritten (1, 2, 3, ..., 100) SELECT kurzIn 10er-Schritten (10, 20, ..., 100) oder SELECTref 20gedrückt halten5. Gewählte Untermessungen ENTERbestätigen und automatische+ 0.0 g *Tierwägung starten.Die Anzahl der Untermessungenbleibt netzausfallsicher gespeichert,bis die Einstellung geändert wird.6. Erstes Tier in den Behälter 888setzen. Die Waage verzögert den 20Beginn der Tierwägung so lange, 19bis die Abweichung von ....2 Messungen das Kriterium erfüllt. 1$ Bei manuellem StartENTER7. Resultat ablesen.Das Resultat der Wägung wird mit+ 410.1 g G*Symbol »*« (= errechneter Wert) mDef 20angezeigt und bleibt so lange in x-Net + 410.1 gder Anzeige stehen, bis das Wägegutvonder Waagschale oderaus dem Behälter entfernt wird.8. Waage entlasten + 0.0 g *9. Ggf. nächstes Tier wägenNächste Messung startet automatisch.91


Netto-TotalAnzeigensymbol: RZweckMit diesem Anwendungsprogrammkönnen Komponenten für einGemisch eingewogen werden.Total- und Komponentengewichtekönnen ausgedruckt werden.Vorbereitung§ Anwendungsprogramm »Netto-Total« im Menü einstellen:siehe Kapitel »Voreinstellungen«§ Parameter einstellen:Anwend.ungsprogrammenet.tot. Netto-TotalKo.Druck KomponentendruckAuso EinMerkmale– Einwaage einzelner Komponenten(maximal 99 Komponenten) vomAnzeigewert »Null« bis zum gewünschtenKomponentengewicht.– Übernehmen der eingewogenenKomponente »Comp xx« mit– anschließender Nullstellung derAnzeige und– automatischem Ausdruck– Löschen des Komponentenspeichersnach Abbruch der Messreihemit Taste CLEAR und Ausdruck desGesamtgewichts.– Umschalten zwischen Komponentengewichtund Gesamtgewichtmit Taste SELECT .– Ausdruck der Summe der Komponentengewichte(T-Comp)o = WerkseinstellungAusdruck Netto-TotalComp 2+ 278.1 g : Zweites KomponentengewichtT-Comp+ 2117.5 g : Summe derKomponenten92


Beispiel: Mehrere Komponenten einwiegenVoreinstellungen für Parameter:Anwend. - Net.tot. (Code 2. 5.)Schritt Taste drücken Anzeige/Ausgabe1. Leeren Behälter zum Einfüllender Komponenten auf die .Waage stellen65.0 g2. Tarieren t 0.0 g3. Erste Komponente einfüllen + 120.5 g4. Komponente übernehmen ENTER+ 0.0 g *NETComp 1+ 120.5 g5. Nächste Komponente einfüllen + 70.5 g *NET6. Komponente übernehmen ENTER+ 0.0 g *NET7. Ggf. weitere Komponenten Schritt 5 und 6einfüllenentsprechend oftwiederholenComp 2+8. Ggf. zum gewünschten Endwert SELECTauffüllen+ 191.0 g *(Umschalten in Totalanzeige)70.5 g9. Gesamtgewicht ausdrucken und CLEAR+ 2117.5 gKomponentenspeicher löschen T-Comp+ 2117.5 g93


SummierenAnzeigensymbol: RZweckMit diesem Anwendungsprogrammkönnen Wägewerte voneinanderunabhängiger Wägungen in einemSpeicher summiert werden, die denWägebereich der Waage überschreiten.Vorbereitung§ Anwendungsprogramm»Summieren« im Menü einstellen:siehe Kapitel »Voreinstellungen«§ Parameter einstellen:Anwend.ungsprogrammesumme SummierenKo.Druck KomponentendruckAuso Eino = WerkseinstellungMerkmale– Summenspeicher für bis zu99 Posten– Übernehmen der eingewogenenKomponente »Comp xx« mit automatischemAusdruck– Umschalten zwischen der Anzeigeaktueller Wägewert und dem Wertim Summenspeicher mit Taste SELECT .– Ausdruck der Summe der Komponentengewichte(S-Comp)– Anwendung beenden und das Summengewichtausdrucken: Taste CLEARdrückenAusdruck SummierenComp 2+ 278.1 g : ZweitesKomponentengewichtS-Comp+ 2117.5 g : Summenspeicher94


Beispiel: Wägewerte summierenVoreinstellungen für Parameter:Anwend. - summe - ko.druck - ein (Code 2. 6. 1. 2)Schritt Taste drücken Anzeige/Ausgabe1. Tarieren t 0.0 g2. Wägegut auf die Waage legen + 380.0 g(hier z.B. 380 g)3. Wert in Speicher übernehmen ENTER+ 380.0 g *Comp 1+4. Wägegut herunternehmen + 0.0 g *380.0 g5. Nächstes Wägegut auf dieWaage legen (hier z.B. 575 g) + 575.0 g *6. Wert in Speicher übernehmen ENTER+ 955.0 g *+ 575.0 gComp 2+ *575.0 g7. Wert im Summenspeicher SELECT+ 955.0 g Gansehen*8. Ggf. weitere Komponenten Schritt 5 und 6übernehmenentsprechend oftwiederholen9. Gesamtgewicht ausdrucken und CLEAR0.0 gSummenspeicher löschen S-Comp+ 2117.5 g95


DichtebestimmungAnzeigensymbol: WZweckMit diesem Anwendungsprogrammkann die Dichte für feste Stoffe nachder Auftriebsmethode bestimmtwerden. Das Ergebnis kann mit eineroder ohne Nachkommastelle angezeigtwerden: siehe Kapitel »Voreinstellungen«.Nicht im Lieferumfang:Korb und Draht.Ausdruck DichtebestimmungWa + 20.0 g : Wägewert in LuftWfl + 15.0 g : Wägewert inFlüssigkeitRho 4.0 o : Ergebnis: Dichtedes WägegutesVorbereitung§ Anwendungsprogramm »Dichtebestimmung«im Menü einstellen:siehe Kapitel »Voreinstellungen«§ Parameter einstellen:Anwend.ungsprogrammedichte Dichtebestimmungn.komma. Nachkommastellenohne Nachkommastelleo 1nks. 1 Nachkommastelleo = Werkseinstellung96


Beispiel: Dichte eines festen Wägegutes bestimmen.Voreinstellungen für Parameter:Anwend. - Dichte - N.komma -1nks. (Code 2. 9. 1. 2)Schritt Taste drücken Anzeige/Ausgabe1. Korb mit Draht montieren2. Waage tarieren t 0.0 g3. Anwendung starten4. Anzeige »luft« bestätigenENTERENTERluft ?5. Gewicht des Wägegutes in Luftbestimmen:Wägegut auf die Waage legen + 20.0 g *6. Wägewert in Luft übernehmen ENTER7. Wägegut von der Waage nehmen wasser ?8. Wägewert in Flüssigkeit:bestimmenWägegut in den Korb legen.9. Anzeige »water« bestätigen ENTER0.0 g *10. Wägegut in Flüssigkeit tauchen + 15.0 g *11. Wägewert in FlüssigkeitENTERübernehmen, Ergebnis anzeigen + 4.0 ound ausdrucken Wa +*20.0 gWfl + 15.0 gRho4.0 o12. Ergebnis löschen13. Ggf. weiter bei 3.CLEAR97


EinheitenwechselZweckMit diesem Anwendungsprogrammkann ein Wägewert in einer Basiseinheitund in bis zu 4 applikativen Einheitenangezeigt werden (siehe Tabelleauf der nächsten Seite).Vorbereitung§ Anwendungsprogramm »Einheitenwechsel«im Menü einstellen:siehe Kapitel »Voreinstellungen«§ Parameter einstellen:Anwend.ungsprogrammeEinheit EinheitenwechselStellen Anzeigegenauigkeito alle Alle Stellenminus1 Reduziert um1 StelleMerkmale– Die Basiseinheit und deren Anzeigegenauigkeitwird im Menü eingestellt:siehe Kapitel »Voreinstellungen«.– Im Anwendungs-Menü erfolgt nurdie Einstellung der AnwendungEinheitenwechsel und der Anzeigegenauigkeitfür die applikativeEinheiten.– Die Auswahl der Einheiten wirdnetzausfallsicher gespeichert.– Nach dem Einschalten startet dieWaage immer mit der gewähltenBasiseinheit.o = WerkseinstellungAusdruck Einheitenwechsel+ 100.0 g : Wägewert mit Datenausgabeformat 16. ZeichenN + 0.22046 lb : Wägewert mit Datenausgabeformat 22. Zeichen13-Jan-2005 08:35 : Datenausgabeformat für Zweizeilendruck:N + 3.5275 ozt : Datum/Uhrzeit und Wägewert98


Beispiel: Einheit wechseln von Gramm [g] (Basiseinheit) nach Pound [lb] und Troyounce [ozt].Voreinstellung: Anwend. - einheit (Code 2. 2.)Schritt Taste drücken AnzeigeVorbereitung:1. Auswahl für die applikativeGewichtseinheit starten SELECTKEINE o2. Applikative Einheit wählen,hier »Pound«mehrmals(siehe Tabelle: nächste Seite) SELECTpound3. Gewichtseinheit Pound bestätigen ENTERpound o4. Nächste applikative Einheit ENTER , KEINE owählen, hier »Troy ounce« mehrmals(siehe Tabelle: nächste Seite) SELECTtroy.oz.5. Gewichtseinheit »Troy ounce«bestätigen6. Ggf. weitere max. 4 Einheitenkönnen gewählt werden (sonst»keine« mit ENTER bestätigen)troy.oz. o7. Auswahl speichern CLEAR0.00 gBetrieb:8. Wägegut auflegen + 100.00 gENTER9. Gewichtseinheit wechseln mehrmals + 0.22046 lbENTER+ 3.5275 ozt99


Die Waage kann je nach Anforderung mit den folgenden Einheiten arbeiten (beigeeichten Waagen nur möglich, falls das nationale Eichgesetz dies zulässt):Menüpunkt Einheit Umrechnung Anzeigesymbol1) Freie Einheit 1 ) Gramm 1,00000000000 o2) Gramm (Werkseinstellung) Gramm 1,00000000000 g3) Kilogr. Kilogramm 0,00100000000 kg4) carat Carat 5,00000000000 o5) Pound Pound 0,00220462260 lb6) ounce Ounce 0,03527396200 oz7) troy.oz. Troy ounce 0,03215074700 ozt8) tl.honk. Tael Hongkong 0,02671725000 tl9) tl.sing. Tael Singapur 0,02645544638 tl10) tl.taiw. Tael Taiwan 0,02666666000 tl11) grain Grain 15,4323583500 GN12) pennyw. Pennyweight 0,64301493100 dwt13) milligr. Milligramm 1000,00000000 mg14) part./pd. Parts pro pound 1,12876677120 o15) tl.china Tael China 0,02645547175 tl16) momme Momme 0.26670000000 m17) Karat Karat 5,00000000000 Kt18) tola Tola 0.08573333810 o19) baht Baht 0.06578947436 b20) mesghal Mesghal 0.21700000000 o21) tonne Tonne 0,00000100000 t22) pd./oz. 2 ) Pound : Ounces 0,03527396200 lb oz(lb/oz)23) newton Newton 0.00980665000 N1 ) = Vom Anwender definierte Einheit mit PC-Programm in die Waage ladbarüber optionale Schnittstelle RS-232 oder USB.2) = Das Format für Pound:Ounces ist fest realisiert mit xx:yy.yyy; x=lb, y=oz! Bei geeichten Waagen können Gewichtseinheiten gesperrt sein, je nachnationaler Eichgesetzgebung.100


ISO/GLP-ProtokollMerkmaleDie Gerätedaten und Identnummernsowie aktuelles Datum können vor(GLP-Kopf) und nach den Werten derMessreihe (GLP-Fuß) ausgedrucktwerden. Es sind dies:GLP-Kopf:– Datum– Uhrzeit bei Beginn der Messreihe– Waagenhersteller– Waagenmodell– Seriennummer des Modells– Software Versionsnummer– Identifikationsnummer der MessreiheGLP-Fuß:– Datum– Uhrzeit bei Ende der Messreihe– Unterschriftsfeld! Waage mit Drucker betreiben(z.B. <strong>Denver</strong> Instr. GmbH, Drucker901042.1):§ Folgende Einstellungen an derWaage und dem Druckervornehmen:– Software Handshake:setup - interf. - handshk. - softw.(Code 1. 5. 4. 1)Voreinstellung§ Folgende Menüpunkte einstellen(Einstellmodus siehe Kapitel»Voreinstellungen«):– ISO/GLP-konforme Protokollierungnur bei Kalibrieren/Justieren:setup - protok. - Glp - cal.Just.(Code 1. 6. 7. 2)oderISO/GLP-konforme Protokollierungimmer an: setup - protok. - Glp -immer (Code 1. 6. 7. 3)– Zeilenformat des Ausdrucks mitKennzeichnung – 22 Zeichen(Werksvoreinstellung):setup - protok. - format - 22 zei.(Code 1. 6. 6. 2)– Uhrzeit ausgeben im folgendenFormat:setup - protok. - Uhrzeit - 24h(Code 1. 6. 8. 1)odersetup - protok. - Uhrzeit -12hmit »Ahh:mm« oder »Phh:mm«(Code 1. 6. 8. 2)– Datum ausgeben im folgendemFormat:setup - protok. - Datum - tt.mmm.jj(Code 1. 6. 9. 1)odersetup - protok. - DAtum - mmm.tt.jj(Code 1. 6. 9. 2)! Bei folgenden Einstellungenwerden keine ISO/GLP-Protokolleausgegeben: setup - protok. -druck - auto.ohn. oder auto.mit.(Code 1. 6. 1. 3, 1. 6. 1. 4) undformat -16 zei. (Code 1. 6. 6. 1)FunktionstastenProtokollkopf und ersten Messwertausgeben: Taste PRINT drücken> Mit 1. Druck wird Protokollkopfausgegeben.Protokollkopf und Referenzdaten mitautomatischem Druck ausgeben beiaktiviertem Anwendungsprogramm:Taste ENTER drückenAnwendung beenden:1) Protokollfuß ausgeben:Taste CLEAR2) Anwendungsprogramm beenden:Taste CLEAR101


Das ISO/GLP-Protokoll kann aus folgenden Zeilen bestehen:-------------------- Strichzeile17-Aug-2005 10:15 Datum/Uhrzeit (Beginn der Messung)DENVERWaagenherstellerMod. SI-8001 WaagentypSer. no. 10105355 Seriennummer der WaageVer. no. 00-36-01 Software-VersionID 2690 923 ID.-------------------- StrichzeileL IDMessreihe-Nr.nRef 10 pcs Zählen: ReferenzstückzahlwRef 21.14 g Zählen: ReferenzgewichtQnt + 567 pcs Zählergebnis-------------------- Strichzeile17-Aug-2005 10:20 Datum/Uhrzeit (Ende der Messung)Name:UnterschriftsfeldLeerzeile-------------------- StrichzeileDas ISO/GLP-Protokoll für externes Kalibrieren/Justieren:-------------------- Strichzeile17-Aug-2005 10:30 Datum/Uhrzeit (Beginn der Messung)DENVERWaagenherstellerMod. SI-8001 WaagentypSer. no. 10105352 Seriennummer der WaageVer. no. 00-36-01 Software-VersionID 2690 923 Ident-Nr.-------------------- StrichzeileCal. Ext. Test Art des Kalibrierens/JustierensSet + 5000.0 g JustiergewichtswertDiff. + 0.2 g Differenz nach KalibrierungCal. Ext. Complete Bestätigung des abgeschlossenen JustiervorgangsDiff. 0.0 g Differenz zum Sollwert nach Justierung-------------------- Strichzeile17-Aug-2005 10:32 Datum/Uhrzeit (Ende der Messung)Name:UnterschriftsfeldLeerzeile-------------------- Strichzeile102


DatenschnittstelleZweckDie Waage besitzt eine Datenschnittstelle,an die ein Rechner (oder einanderes Peripheriegerät) angeschlossenwerden kann.Mit einem Rechner können Waagenfunktionenund Funktionen derAnwendungsprogramme verändert,gestartet und überwacht werden.VorbereitungDie Anpassung an das angeschlosseneGerät im Menü vornehmen:siehe Kapitel »Voreinstellungen«.Eine detaillierte Beschreibung derverfügbaren Schnittstellenbefehle,Kommandos und Datenausgangsformatefordern Sie bitte direkt bei<strong>Denver</strong> an.Die vielseitigen Eigenschaften derWaagen bezüglich Dokumentationder Resultate lassen sich erst mit demAnschluss eines Druckers von <strong>Denver</strong>voll nutzen. Die Druckresultate tragenzu einer einfachen Arbeitsweise nachGLP entscheidend bei.Steckerbelegungsplan (Handshake)Mit Hilfe einer 4-poligen Schnittstelle können 1 oder 2 weitere Zeichen nachdem CTS-Signal übertragen werden.Diese Verbindungen müssen hergestellt werden, wenn die Waage über denRS232C-Port angeschlossen werden soll!Pinbelegung des Daten-Schnittstellenkabels für Waagen der <strong>Summit</strong>-Serie:<strong>Summit</strong>-SerieStandard RS2329- poligen Port der Waage Anschlussstecker, 9- poligRxD 23 TxDTxD 32 RxDDTR 44 DTRSignalerdung 55 SignalerdungCTS 88 CTS<strong>Summit</strong>-SerieStandard RS2329- poliger Port der Waage Anschlussstecker, 25- poligRxD 22 TxDTxD 33 RxDDTR 420 DTRSignalerdung 57 SignalerdungCTS 85 CTS103


FehlermeldungenFehlermeldungen werden in der Hauptanzeige für ca. 2 Sekunden dargestellt.Danach kehrt das Programm automatisch in den Wägezustand zurück.Anzeige Ursache Abhilfekeine Anzeigesegmente Keine Betriebsspannung StromversorgungerscheinenüberprüfenNetzgerät nichtNetzgerät an die Stromeingestecktversorgung anschließenhigh Wägebereich überschritten Waagschale entlastenlow oder Err 54 Berührung zwischen Waagschale darf umge-Lastplatte und Umgebung bende Teile nicht berührenerr 54, typisch Wägesystem defekt <strong>Denver</strong> KundendienstansprechenApp.err. Übernahmefehler: Gewicht erhöhenGewicht zu gering oderkein Wägegut auf derWaagschale bei Anwendungsprogrammendis.err. Datenausgabe passt nicht Korrekte Einstellung imins AusgabeformatMenü vornehmenprt.err. Datenschnittstelle für Menü-Reset durchführenDruckausgabe gesperrt oder<strong>Denver</strong>-Kundendienstansprechenerr 02 Justierbedingung wurde Erst nach Nullanzeigenicht eingehalten, z.B.: justieren– Tarieren mit Taste t– Waagschale belastet Waage entlastenerr 10 Taste t gesperrt Erst nach Löschen desbei aktiven Anwendungs- Taraspeichers über Tasteprogramm »Netto-Total«; CLEAR ist die Taste tTarafunktionen sind wieder ausführbargegeneinander verriegelterr 11 Taraspeicher nicht erlaubt Taste t drückenWägewert ändert sich Aufstellort instabil Aufstellort wechselnlaufend (zuviel Vibration oder Anpassung im SetupLuftzug vorhanden)vornehmenFremdkörper zwischen Fremdkörper entfernenWaagschale und GehäuseOffensichtlich falsches Waage nicht justiert JustierenWägeergebnis Vor dem Wägen nicht TarierentariertFalls andere Fehler auftreten, <strong>Denver</strong>-Kundendienst anrufen!Adressen: siehe Internet: http://www.denverinstrument.com104


Pflege und WartungServiceAuf Anforderung kann ein Wartungsvertragindividuell vereinbart werden.ReparaturenReparaturen dürfen nur von Fachkräften ausgeführtwerden. Durch unsachgemäße Reparaturenkönnen erhebliche Gefahren für den Benutzerentstehen.Reinigung§ Waage von der Spannungsversorgungtrennen, ggf. angeschlossenes Datenkabelan der Waage lösen.! Es darf keine Flüssigkeit in die Waagegelangen.! Keine aggressiven Reinigungsmittelverwenden (Lösungsmittel o.ä.).§ Waage mit weichem Tuch abtrocknen.Waagschale bei Analysenwaagen abnehmenund reinigen:§ Unter den Schirmring fassen und zusammenmit der Unterschale die Waagschale nachoben führen, damit das Wägesystem nichtbeschädigt wird.! Es darf keine Flüssigkeit in die Waagegelangen.Reinigung der EdelstahloberflächenGrundsätzlich alle Edelstahlteile in regelmäßigenAbständen reinigen. Edelstahllastplatte separatgründlich reinigen. Edelstahlteile an der Waagemit einem feuchten Tuch oder Schwammreinigen. Nur handelsübliche Haushaltsreinigerverwenden, die für Edelstahl geeignet sind(z.B. Stahlfix). Edelstahloberflächen durcheinfaches Abreiben reinigen. Danach gründlichnachspülen, bis alle Rückstände beseitigt sind.Anschließend das Gerät trocknen lassen.Als zusätzlicher Schutz kann ein Pflegeöl aufgetragenwerden.105


EntsorgungSicherheitsüberprüfungErscheint ein gefahrloser Betrieb derWaage nicht mehr gewährleistet:§ Spannungsversorgung trennen:Netzgerät aus der Steckdose ziehen.> Waage vor weiterer Benutzungsichern<strong>Denver</strong>-Kundendienst benachrichtigen.Instandsetzungsmaßnahmen dürfenausschließlich von Fachkräften ausgeführtwerden.Eine regelmäßige Überprüfung durcheinen Fachmann wird für folgendePunkte empfohlen:– Isolationswiderstand > 7 MOhm miteiner Gleichspannung von mindestens500 V bei 500 kOhm Last– Ersatz-Ableitstrom < 0.05 mA miteinem bestimmungsgemäßenMessgerätEntsorgungWird die Verpackung nicht mehr benötigt,ist diese der örtlichen Müllentsorgungzuzuführen. Die Verpackungbesteht aus umweltfreundlichen Materialien,die als Sekundärrohstoffe dienenkönnen.Das Gerät inklusive Zubehörund Batterien gehört nichtin den Hausmüll. Die EU-Gesetzgebung fordert inihren Mitgliedsstaaten,elektrische und elektronischeGeräte vom unsortierten Siedlungsabfallgetrennt zu erfassen, um sieanschließend wiederzuverwerten.In Deutschland und einigen anderenLändern führt die <strong>Denver</strong> <strong>Instrument</strong>die Rücknahme und gesetzeskonformeEntsorgung ihrer elektrischen undelektronischen Produkte selbst durch.Diese Produkte dürfen nicht – auchnicht von Kleingewerbetreibenden –in den Hausmüll oder an Sammelstellender örtlichen öffentlichen Entsorgungsbetriebeabgegeben werden.Hinsichtlich der Entsorgung wendenSie sich daher in Deutschland wieauch in den Mitgliedsstaaten desEuropäischen Wirtschaftsraumesbitte an unsere Service-Mitarbeitervor Ort oder an unsere Service-Zentrale:<strong>Denver</strong> <strong>Instrument</strong> GmbHRobert-Bosch-Breite 1037079 Göttingen, DeutschlandIn Ländern, die keine Mitglieder desEuropäischen Wirtschaftsraumes sindoder in denen es keine <strong>Denver</strong>-Filialengibt, sprechen Sie bitte die örtlichenBehörden oder Ihr Entsorgungsunternehmenan.Vor der Entsorgung bzw. Verschrottungdes Gerätes sollten die Batterienentfernt werden und einer Sammelstelleübergeben werden.Mit gefährlichen Stoffen kontaminierteGeräte (ABC-Kontamination) werdenweder zur Reparatur noch zur Entsorgungzurückgenommen. AusführlicheInformationen mit Service-Adressenzur Reparaturannahme oder EntsorgungIhres Gerätes können Sie aufunserer Internetseite (www.denverinstrument.com)finden oder über den<strong>Denver</strong> Service anfordern.106


ÜbersichtTechnische DatenAllgemeine technische DatenInt. Justiergewichtsschaltung Alle SI-ModelleNetzanschluss, über Steckernetzgerät 230 V oder 115 V,-spannung, -frequenz+15%…– 20%, 48–60 HzLeistungsaufnahme VA maximal 16; typisch 8(einschl. Netzgerät STNG6)Betriebsdauer mitexternem Akku SBP01(Hinterleuchtung ein), ca. h 35107


Modellspezifische technische DatenModell SI-234 | S-234 SI-114 | S-114 SI-64 | S-64Wägebereich g 230 110 60Ablesbarkeit g 0,0001 0,0001 0,0001Tarierbereich (subtraktiv) g 230 110 60Reproduzierbarkeit


Modellspezifische technische DatenModell SI-6002| SI-4002| SI-2002|S-6002 S-4002 S-2002Wägebereich g 6000 4000 2000Ablesbarkeit g 0,01 0,01 0,01Tarierbereich (subtraktiv) g 6000 4000 2000Reproduzierbarkeit


Geeichte Modelle mit EG-Bauartzulassung: Modellspezifische technische DatenModell SI-234A SI-114A SI-64ATypbezeichnung BD ED 100 BD ED 100 BD ED 100Genauigkeitsklasse 1 ) k k kWägebereich Max. 1 ) g 230 110 60Ziffernschritt d 1 ) g 0,0001 0,0001 0,001Tarierbereich (subtraktiv)< 100% vom maximalen WägebereichEichwert e 1 ) g 0,001 0,001 0,001Mindestlast Min. 1 ) g 0,01 0,01 0,01Einschwingzeit (typisch) s 2,5Verwendungsbereichnach RL 1 ) g 0,01 – 230 0,01 – 110 0,01 – 60Temperatur °C +17…+27Nettogewicht, ca. kg 4,8Waagschalenabmessung mm 90 dWägeraumhöhe mm 230Abmessungen (B + T + H) mm 230 + 303 + 330Modell SI-603A SI-403A SI-203ATypbezeichnung BD ED 200 BD ED 200 BD ED 200Genauigkeitsklasse 1 ) K K KWägebereich Max. 1 ) g 600 400 200Ziffernschritt d 1 ) g 0,001 0,001 0,001Tarierbereich (subtraktiv)< 100% vom maximalen WägebereichEichwert e 1 ) g 0,01 0,01 0,01Mindestlast Min. 1 ) g 0,02 0,02 0,02Einschwingzeit (typisch) s 1 1 1Verwendungsbereichnach RL 1 ) g 0,02 – 600 0,02 – 400 0,02 – 200Temperatur °C +10…+30Nettogewicht, ca. kg 3,6Waagschalenabmessung mm 115 dAbmessungen (B + T + H) mm 230 + 303 + 1361) Richtlinie 90/384/EWG für nichtselbsttätige Waagen für den Bereichdes Europäischen Wirtschaftsraumes110


Geeichte Modelle mit EG-Bauartzulassung: Modellspezifische technische DatenModell SI-6002A SI-4002A SI-2002ATypbezeichnung BD ED 200 BD ED 200 BD ED 200Genauigkeitsklasse 1 ) K K KWägebereich Max. 1 ) g 6000 4000 2000Ziffernschritt d 1 ) g 0,01 0,01 0,01Tarierbereich (subtraktiv)< 100% vom maximalen WägebereichEichwert e 1 ) g 0,1 0,1 0,1Mindestlast Min. 1 ) g 0,5 0,5 0,5Einschwingzeit (typisch) s 1,1 1,1 1,1Verwendungsbereichnach RL 1 ) g 0,5 – 6000 0,5 – 4000 0,5 – 2000Temperatur °C +10…+30Nettogewicht, ca. kg 3,5Waagschalenabmessung mm 180 + 180Abmessungen (B + T + H) mm 230 + 303 + 91Modell SI-602A SI-8001A SI-4001ATypbezeichnung BD ED 200 BD ED 200 BD ED 200Genauigkeitsklasse 1 ) K K KWägebereich Max. 1 ) g 600 8000 4000Ziffernschritt d 1 ) g 0,01 0,1 0,1Tarierbereich (subtraktiv)< 100% vom maximalen WägebereichEichwert e 1 ) g 0,1 1 1Mindestlast Min. 1 ) g 0,5 5 5Einschwingzeit (typisch) s 1,1 1 1Verwendungsbereichnach RL 1 ) g 0,5 – 600 5 – 8000 5 – 4000Temperatur °C +10…+30Nettogewicht, ca. kg 3,5Waagschalenabmessung mm 180 + 180Abmessungen (B + T + H) mm 230 + 303 + 911) CD = Council Directive 90/384/EEC for non-automatic weighing instruments;applicable to the European Economic Area111


ZubehörArtikelWindschutzring für Modelle mit runder WaagschaleArbeitsschutzhaube für BedienfeldBestell-Nr.auf Anfrageauf AnfrageDrucker mit Verbindungskabel 901042.19-poliges Kabel für die Waage 400191-1DiebstahlsicherungUnterflurwägeeinrichtung*DichtebestimmungssetExterner Akkuauf AnfrageYDK01LPSBP01Justiergewichte – für eine vollständige Liste wendenSie sich bitte an Ihren Händler* = Für eichpflichtige Anwendungen nicht erlaubt112


C – KennzeichnungDas Gerät erfüllt die Anforderungender Richtlinien des Rates derEuropäischen Union:89/336/EWG »ElektromagnetischeVerträglichkeit (EMV)«Zugehörige Europäische Normen:Störaussendung:nach Produktnorm EN 61326-1 Kl.B(Wohnbereich)Störfestigkeit:nach Produktnorm EN 61326-1(Mindestprüfanforderungen, nichtkontinuierlicher Betrieb)Hinweis:Modifikationen der Geräte sowieder Anschluss von nicht von <strong>Denver</strong>gelieferten Kabeln oder Gerätenunterliegen der Verantwortung desBetreibers und sind von diesementsprechend zu prüfen und fallserforderlich zu korrigieren.<strong>Denver</strong> stellt auf Anfrage Angabenzur Betriebsqualität zur Verfügung(gemäß den o.g. Normen zurStörfestigkeit).73/23/EWG »Elektrische Betriebsmittelzur Verwendung innerhalbbestimmter Spannungsgrenzen«Zugehörige Europäische Normen:EN 60950Sicherheit von Einrichtungen derInformationstechnik, einschließlichelektrischer BüromaschinenEN61010Sicherheitsanforderungen anelektrische Mess-, Steuer-, RegelundLaborgeräteTeil 1: Allgemeine AnforderungenBei Verwendung elektrischerBetriebsmittel in Anlagen und Umgebungsbedingungenmit erhöhtenSicherheitsanforderungen sind dieAuflagen gemäß den zutreffendenErrichtungsbestimmungen zubeachten.113


114


Copyright by <strong>Denver</strong> <strong>Instrument</strong> GmbH, Goettingen, Germany.All rights reserved. No part of this publication may be reprintedor translated in any form or by any means without the priorwritten permission of <strong>Denver</strong> <strong>Instrument</strong> GmbH.The status of the information, specifications and illustrations inthis manual is indicated by the date given below. <strong>Denver</strong> <strong>Instrument</strong>GmbH reserves the right to make changes to the technology, features,specifications, and design of the equipment without notice.Status: January 2007, <strong>Denver</strong> <strong>Instrument</strong> GmbH, Goettingen,Germany.Europe, Asia and Australia:Internet:<strong>Denver</strong> <strong>Instrument</strong> GmbHRobert-Bosch-Breite 1037079 Goettingen, GermanyTel: +49-551-20977-30Fax: +49-551-20977-39www.denverinstrument.comU.K. and Ireland:North and South America:<strong>Denver</strong> <strong>Instrument</strong> Company<strong>Denver</strong> House, Sovereign WayTrafalgar Business ParkDownham MarketNorfolk PE38 9SW EnglandTel: +44-136-63862-42Fax: +44-136-63862-04<strong>Denver</strong> <strong>Instrument</strong> Company1401 17 th Street, Suite 750<strong>Denver</strong>, Colorado 802021-800-321-1135Tel: +1-303-431-7255Fax: +1-303-423-4831W4A000 · KTPublication No. WSI6004-a0701698648-013-93